Background: Sacral screw loosening is a typical complication after internal fixation surgery through the vertebral arch system. Bicortical fixation can successfully prevent screw loosening, and how improving the rate of bicortical fixation is a challenging clinical investigation.
Objective: To investigate the feasibility of improving the double corticality of sacral screws and the optimal fixation depth to achieve double cortical fixation by combining the torque measurement method with bare hands.
Methods: Ninety-seven cases of posterior lumbar internal fixation with pedicle root system were included in this study. Based on the tactile feedback of the surgeon indicating the expected penetration of the screw into the contralateral cortex of the sacrum, the screws were further rotated by 180°, 360°, or 720°, categorized into the bicortical 180° group, bicortical 360° group, and bicortical 720° group, respectively. Intraoperatively, the torque during screw insertion was recorded. Postoperatively, the rate of double-cortex engagement was evaluated at 7 days, and screw loosening was assessed at 1 year follow-up.
Results: The bicortical rates of the 180° group, 360° group, and 720° group were 66.13%, 91.18% and 93.75%, respectively. There were statistically significant differences between the 180° group and both the 360° and 720° groups (P < 0.05). However, there was no statistically significant difference between the 360° group and the 720° group (P > 0.05).The rates of loosening of sacral screws in the 180° group, 360° group, and 720° group were 20.97%, 7.35% and 7.81%, respectively. There were statistically significant differences between the 180° group and both the 360° and 720° groups (P < 0.05). However, there was no statistically significant difference between the 360° group and the 720° group (P > 0.05). The bicortical 360° group achieved a relatively satisfactory rate of dual cortical purchase while maintaining a lower rate of screw loosening.
Conclusion: Manual insertion of sacral screws with the assistance of a torque measurement device can achieve a relatively satisfactory dual cortical purchase rate while reducing patient hospitalization costs.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11225279 | PMC |
http://dx.doi.org/10.1186/s12891-024-07627-3 | DOI Listing |