Purpose: This study aimed to develop and validate a Monte Carlo (MC) model for the Papillon+ contact x-ray brachytherapy (CXB) device, producing 50 kilovolt (kV) X-rays, specifically focusing on its application with a 25 mm diameter rectal applicator for contact therapy.

Material And Methods: The validation process involved depth dose and transverse dose profile measurements using EBT3 gafchromic films positioned in a plastic water low energy range phantom. The half-value layer (HVL) was further measured and derived from the simulated X-ray spectra.

Results: Excellent agreement within ±2% was achieved between the measured and simulated on-axis depth dose curves for the 25 mm rectal applicator. Transverse dose profile measurements showed a high level of agreement between the simulation and measurements, on average 3.1% in contact with the applicator at the surface of the phantom and on average 1.7% at 10 mm depth. A close agreement within 5.5% was noticed concerning the HVL between the measurement and simulation. The simulated gamma spectra and 2D-dose distribution demonstrated a soft X-ray energy spectrum and a uniform dose distribution in contact with the applicator.

Conclusions: An MC model was successfully developed for the Papillon+ eBT device with a 25 mm diameter rectal applicator. The validated model, with its demonstrated accuracy in depth dose and transverse dose profile simulations, is a valuable tool for quality assurance and patient safety and, in a later phase, may be used for treatment planning, dose calculations and tissue inhomogeneity corrections.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.brachy.2024.06.003DOI Listing

Publication Analysis

Top Keywords

rectal applicator
12
depth dose
12
transverse dose
12
dose profile
12
monte carlo
8
papillon+ contact
8
contact x-ray
8
x-ray brachytherapy
8
diameter rectal
8
dose
8

Similar Publications

Purpose: To present an innovative deformable applicator that used Freiburg flap as vaginal applicator with or without free-hand interstitial needles in three-dimensional (3D) high-dose-rate (HDR) brachytherapy for vaginal stump recurrence of cervical cancer.

Material And Methods: Between September 2017 and January 2020, all patients with vaginal stump recurrence after radical hysterectomy of cervical cancer treated with vaginal stump brachytherapy using Freiburg flap as vaginal applicator with or without free-hand interstitial needles were retrospective analyzed. Characteristics related to patients and treatment modality as well as preliminary outcomes and side effects were investigated.

View Article and Find Full Text PDF

Purpose: The bladder and rectal toxicities in cervical cancer brachytherapy are positively correlated with the DVH parameter: D2cc. This study evaluates the feasibility of knowledge-based planning to predict the D2cc, identify suboptimal plans, and improve the plan quality with Direction Modulated Brachytherapy (DMBT) applicators using knowledge-based planning based on linear relationship between overlap distances and D2cc.

Methods: The overlap volume histogram (OVH) method was used to determine the distances for 2 cm of overlap between the Organs at Risks (OAR) and High-Risk Clinical Target Volume (CTV).

View Article and Find Full Text PDF
Article Synopsis
  • The study aimed to compare the effectiveness and safety of different brachytherapy planning methods for treating locally advanced cervical cancer, specifically between CT and MRI-based approaches.
  • A total of 106 patients were analyzed, with results indicating no significant difference in survival rates between the two methods, but MRI-based planning with the Template Benidorm resulted in smaller treatment volumes and reduced toxicity.
  • The findings suggested that MRI planning allows for better treatment precision, leading to decreased rectal hemorrhage and a trend towards higher doses delivered to the tumor site without increasing overall toxicity.
View Article and Find Full Text PDF

Budesonide Rectal Foam (BF) was introduced in 2017 and changed in November 2022 upon request, addressing the challenges encountered with liquid rectal formulations indicated for ulcerative colitis (UC). This formulation is an important agent in the treatment of rectal to sigmoid colon lesions in moderate UC. As the characteristics of the formulation of the rectal formulation are thought to influence patient satisfaction, a survey was conducted on the formulation and patient satisfaction among patients who used BF before and after the change.

View Article and Find Full Text PDF

Introduction: At our department we have a dedicated 1.5 Tesla MRI/HDR brachytherapy suite, which provides the possibility of repeated MRI scanning before, during and after applicator insertion and before and/or after irradiation for patients with advanced cervical cancer. In this study we analysed the effect of this adaptive workflow.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!