A Mueller matrix polarimetry system at 532 nm wavelength is developed for noninvasive glucose sensing in turbid media such as human's fingertip. The system extracts mean absorbance and anisotropic properties, demonstrated numerically and experimentally with phantom glucose samples. It is found that mean absorbance ( ), depolarization index (Δ), and linear dichroism (LD) show linear variation with glucose concentration 100-500 mg/dL. In addition, LightTools simulations indicate proportional scaling of scattering effects with , Δ, and LD. Real-world tests on fingertip show a strong correlation between these properties and blood glucose levels with a mean absolute relative deviation (MARD) of 12.56% and a correlation coefficient (R) of 0.875 in prediction by a neural network (NN) model, highlighting the advantages of Mueller matrix in extracting more parameters related to blood glucose.