A practical method for measuring planar spatial resolution with scatter on a CZT gamma camera.

EJNMMI Phys

Service de Physique Médicale et Radioprotection, Groupement Hospitalier Est, Hospices Civils de Lyon, Bâtiment Pinel, 52 Bd Pinel, Lyon, Bron, 69677, France.

Published: July 2024

Purpose: This study proposes a practical method for evaluating 2D spatial resolution with scatter on a CZT planar detector gamma camera, which is simpler and faster than the NEMA method. It is used to characterize the influence of distance on spatial resolution FWHM on a CZT camera equipped with a WEHR collimator.

Methods: The practical method uses linear sources tilted with respect to the detector axes. The spatial resolution full width at half maximum (FWHM) with four tilt angles was compared to the FWHM evaluated using the NEMA NU1-2018 method. Spatial resolution FWHM was also assessed with tilted sources acquired at distances of 0 to 20 cm using a single angle, with and without the post-processing image enhancement proposed by the manufacturer.

Results: Estimated spatial resolution FWHM with tilted sources was close to the spatial resolution FWHM estimated at 7.63 mm by the NEMA method, with deviations ranging from - 5.62 to 4.59% at 10 cm depending on the angle considered. The study of spatial resolution FWHM dependence on distance indicates that, for distances less than 3 cm, the FWHM no longer decreases with distance. The manufacturer's post-processing reduces the FWHM by an average of 15%.

Conclusion: The practical method is quicker to implement and gives comparable results to the NEMA reference method for spatial resolution FWHM. Evaluation of spatial resolution with linear sources at short distances from the collimator is limited by the collimator effect and signal digitization. The tilted source method can be used to measure spatial resolution quickly and easily under clinical conditions for CZT planar cameras.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11217200PMC
http://dx.doi.org/10.1186/s40658-024-00656-wDOI Listing

Publication Analysis

Top Keywords

spatial resolution
44
resolution fwhm
24
practical method
16
spatial
11
resolution
11
fwhm
10
resolution scatter
8
scatter czt
8
gamma camera
8
method
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!