Background: MR-integrated proton therapy is under development. It consists of the unique challenge of integrating a proton pencil beam scanning (PBS) beam line nozzle with an magnetic resonance imaging (MRI) scanner. The magnetic interaction between these two components is deemed high risk as the MR images can be degraded if there is cross-talk during beam delivery and image acquisition.
Purpose: To create and benchmark a self-consistent proton PBS nozzle model for empowering the next stages of MR-integrated proton therapy development, namely exploring and de-risking complete integrated prototype system designs including magnetic shielding of the PBS nozzle.
Materials And Methods: Magnetic field (COMSOL ) and radiation transport (Geant4) models of a proton PBS nozzle located at OncoRay (Dresden, Germany) were developed according to the manufacturers specifications. Geant4 simulations of the PBS process were performed by using magnetic field data generated by the COMSOL simulations. In total 315 spots were simulated which consisted of a scan pattern with 5 cm spot spacings and for proton energies of 70, 100, 150, 200, and 220 MeV. Analysis of the simulated deflection at the beam isocenter plane was performed to determine the self-consistency of the model. The magnetic fringe field from a sub selection of 24 of the 315 spot simulations were directly compared with high precision magnetometer measurements. These focused on the maximum scanning setting of 20 cm beam deflection as generated from the second scanning magnet in the PBS for a proton beam energy of 220 MeV. Locations along the beam line central axis (CAX) were measured at beam isocenter and downstream of 22, 47, 72, 97, and 122 cm. Horizontal off-axis positions were measured at 22 cm downstream of isocenter ( 50, 100, and 150 cm from CAX).
Results: The proton PBS simulations had good spatial agreement to the theoretical values in all 315 spots examined at the beam line isocenter plane (0-2.9 mm differences or within 1.5 % of the local spot deflection amount). Careful analysis of the experimental measurements were able to isolate the changes in magnetic fields due solely to the scanning magnet contribution, and showed 1.9 1.2 -9.4 1.2 changes over the range of measurement locations. Direct comparison with the equivalent simulations matched within the measurement apparatus and setup uncertainty in all but one measurement point.
Conclusions: For the first time a robust, accurate and self-consistent model of a proton PBS nozzle assembly has been created and successfully benchmarked for the purposes of advancing MR-integrated proton therapy research. The model will enable confidence in further simulation based work on fully integrated designs including MRI scanners and PBS nozzle magnetic shielding in order to de-risk and realize the full potential of MR-integrated proton therapy.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/mp.17279 | DOI Listing |
Med Phys
September 2024
Helmholtz-Zentrum Dresden-Rossendorf, Institute of Radiooncology-OncoRay, Dresden, Germany.
Background: MR-integrated proton therapy is under development. It consists of the unique challenge of integrating a proton pencil beam scanning (PBS) beam line nozzle with an magnetic resonance imaging (MRI) scanner. The magnetic interaction between these two components is deemed high risk as the MR images can be degraded if there is cross-talk during beam delivery and image acquisition.
View Article and Find Full Text PDFPhys Imaging Radiat Oncol
July 2023
Joint Department of Physics at The Institute of Cancer Research and The Royal Marsden NHS Foundation Trust, 15 Cotswold Rd, London SM2 5NG, United Kingdom.
Background And Purpose: Physiological motion impacts the dose delivered to tumours and vital organs in external beam radiotherapy and particularly in particle therapy. The excellent soft-tissue demarcation of 4D magnetic resonance imaging (4D-MRI) could inform on intra-fractional motion, but long image reconstruction times hinder its use in online treatment adaptation. Here we employ techniques from high-performance computing to reduce 4D-MRI reconstruction times below two minutes to facilitate their use in MR-guided radiotherapy.
View Article and Find Full Text PDFMagn Reson Med
November 2023
OncoRay - National Center for Radiation Research in Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität Dresden, Helmholtz-Zentrum Dresden-Rossendorf, Dresden, Germany.
Purpose: The low sensitivity and limitation to water phantoms of convection-dependent MRI magnitude signal-based proton beam visualization hinder its in vivo applicability in MR-integrated proton beam therapy. The purpose of the present study was, therefore, to assess possible contrast mechanisms for MRI phase signal-based proton beam visualization that can potentially be exploited to enhance the sensitivity of the method and extend its applicability to tissue materials.
Methods: To assess whether proton beam-induced magnetic field perturbations, changes in material susceptibility or convection result in detectable changes in the MRI phase signal, water phantom characteristics, experiment timing, and imaging parameters were varied in combined irradiation and imaging experiments using a time-of-flight angiography pulse sequence on a prototype in-beam MRI scanner.
Phys Med Biol
September 2021
Department of Radiation Oncology, Massachusetts General Hospital and Harvard Medical School, Boston, MA, United States of America.
Magnetic resonance imaging (MRI)-integrated proton therapy (MRiPT) is envisioned to improve treatment quality for many cancer patients. However, given the availability of alternative image-guided strategies, its clinical need is yet to be justified. This study aims to compare the expected clinical outcomes of MRiPT with standard of practice cone-beam CT (CBCT)-guided PT, and other MR-guided methods, i.
View Article and Find Full Text PDFRadiat Oncol
May 2020
Department of Medical Physics, Faculty of Physics, Ludwig-Maximilians-Universität München, Garching, Germany.
Background: The targeting accuracy of proton therapy (PT) for moving soft-tissue tumours is expected to greatly improve by real-time magnetic resonance imaging (MRI) guidance. The integration of MRI and PT at the treatment isocenter would offer the opportunity of combining the unparalleled soft-tissue contrast and real-time imaging capabilities of MRI with the most conformal dose distribution and best dose steering capability provided by modern PT. However, hybrid systems for MR-integrated PT (MRiPT) have not been realized so far due to a number of hitherto open technological challenges.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!