Statement Of Problem: Artificial intelligence has been used to enhance the digitalized workflow, especially when undergoing complex oral rehabilitations. However, the reliability of real-time jaw motion registration devices is unclear, and no standard measurement method of the sagittal condylar inclination (SCI) and Bennett angle (BA) has been established.

Purpose: The purpose of this clinical study was to compare and evaluate the reliability of the SCI and BA values recorded by using 2 different digital devices.

Material And Methods: A total of 17 participants, aged between 20 and 30 years (10 women and 7 men) were included in the study. For each participant, the Cadiax Compact 2 and MODJAW tracking devices were used to measure the SCI and BA values at 3 mm and 5 mm of condylar displacement during 3 separate recording sessions. The intraclass correlation coefficient (ICC) was used to assess the reliability of the recordings. Comparisons between the devices were performed with the Wilcoxson rank-sum test (α=.05). The Bland-Altman plot was used to evaluate the quantitative agreement between the 2 devices.

Results: All ICC intrarater reliability values for Cadiax and MODJAW were higher than 0.90. Both at 3 and 5 mm, the SCI measurements were higher for MODJAW registrations than for those recorded by Cadiax. The Bland-Altman plot showed that the SCI values were higher for MODJAW than for Cadiax by 5.9 (95% CI 3.9 to 8.2) and that the BA differences between the MODJAW and the Cadiax were not consistent with the measured value.

Conclusions: Excellent reliability was found with the MODJAW and Cadiax recordings. The SCI and BA values for MODJAW measurements were higher at 3 mm and 5 mm than those acquired with Cadiax. MODJAW showed higher values than Cadiax, and the discrepancies were more pronounced for 3 mm than for 5 mm of condylar displacement, highlighting the need for precision in measurements at lower ranges of motion.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.prosdent.2024.05.014DOI Listing

Publication Analysis

Top Keywords

sci values
16
3 mm 5 mm
12
modjaw cadiax
12
cadiax
9
modjaw
9
cadiax compact
8
compact modjaw
8
condylar inclination
8
5 mm condylar
8
condylar displacement
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!