Objectives: Despite clinical guidelines do not recommend the use of point-of-care testing (POCT) glucometers for diagnostic purposes yet, the analytical performance is continuously improving. Thus, we evaluate the technical accuracy and clinical concordance of POCT glucometers during an oral glucose tolerance test (OGTT) in children for prediabetes and diabetes diagnosis in a comparison study.

Methods: Pediatric patients with an OGTT indication who attended the Diabetes Unit between December 2020 and September 2021 were recruited for this prospective observational study. During the functional test, glycaemia was immediately measured in venous blood using two glucometers (unconnected and connected) and sent to the central laboratory.

Results: The study included 98 patients. There was a high correlation between the glucometers and the central laboratory (Pearson correlation coefficient=0.912 and 0.950, for unconnected and connected glucometer, respectively). The median OGTT turnaround time (TAT) was significantly decreased (connected glucometer: 2.02 h [interquartile range, 2.00-2.07], central laboratory: 11.63 h [6.09-25.80]), with similar overall cost. The diagnostic concordance between connected glucometer and the central laboratory was 71.1 % (95 % confidence interval (CI) 61.5-79.2). The clinical decision would have been the same in the 92.8 % of the cases, but treatment would have not been indicated in 4 patients (4.1 %).

Conclusions: POCT glucometers have demonstrated a high correlation and an acceptable diagnostic concordance with the central laboratory during an OGTT, as well the connected device offers a significant decrease in TAT, without increasing costs. However, as severe clinical impact could happen, POCT glucometers may not be used for diagnosis yet.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11206188PMC
http://dx.doi.org/10.1515/almed-2023-0089DOI Listing

Publication Analysis

Top Keywords

poct glucometers
16
central laboratory
16
connected glucometer
12
glucometers oral
8
oral glucose
8
glucose tolerance
8
tolerance test
8
children prediabetes
8
prediabetes diabetes
8
diabetes diagnosis
8

Similar Publications

A highly sensitive and efficient home-use method is impressive and promising in the self-monitoring of chronic prostatitis (CP). Herein, we developed a glucose oxidase-zeolitic imidazolate framework-8-antibody composite (GOx@ZIF-8@Ab) to achieve highly sensitive point-of-care testing (POCT) of urinary prostatic exosomal protein (PSEP) by combining it with a personal glucometer (PGM). The developed GOx@ZIF-8@Ab was prepared through a simple but effective biomineralization reaction and a strong binding affinity between Zn and Fc region of the antibody.

View Article and Find Full Text PDF

A highly specific and sensitive rapid two-signal assay was developed for the detection of Salmonella typhimurium in foods of animal origin. The invA gene of Salmonella was used as the biorecognition element and recombinase-assisted amplification (RAA) technology for signal amplification. By utilizing the specific recognition and efficient trans-cleavage activity of CRISPR/Cas12a, point-of-care testing (POCT) for S.

View Article and Find Full Text PDF

Objectives: Until now, the external quality assessment (EQA) of glucose point-of-care testing (POCT) has lacked a high quality, suitable and commutable control material to assess measurement accuracy. Here we present a concept for determining the accuracy of glucose measurements, which uses human whole blood and does not require stabilising agents.

Methods: This new generation of quality control samples uses a bead that contains a specific amount of glucose.

View Article and Find Full Text PDF

Objectives: Despite clinical guidelines do not recommend the use of point-of-care testing (POCT) glucometers for diagnostic purposes yet, the analytical performance is continuously improving. Thus, we evaluate the technical accuracy and clinical concordance of POCT glucometers during an oral glucose tolerance test (OGTT) in children for prediabetes and diabetes diagnosis in a comparison study.

Methods: Pediatric patients with an OGTT indication who attended the Diabetes Unit between December 2020 and September 2021 were recruited for this prospective observational study.

View Article and Find Full Text PDF

Sucrose-Powered Liposome Nanosensors for Urinary Glucometer-Based Monitoring of Cancer.

Angew Chem Int Ed Engl

July 2024

State Key Laboratory of Analytical Chemistry for Life Science, College of Engineering and Applied Sciences, Nanjing University, Nanjing, 210023, China.

Timely detection of early-stage cancer holds immense potential in enhancing prognostic outcomes. There is an increasing desire for versatile tools to enable simple, sensitive, and cost-effective cancer detection. By exploiting the extraintestinal metabolic inertness and efficiency renal clearance of sucrose, we designed a liposome nanosensor using sucrose as a messenger to convert tumor-specific esterase activity into glucose meter readout, enabling economical and sensitive urinalysis for cancer detection in point-of-care testing (POCT).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!