Background: To evaluate the agreement between the Goldman applanation tonometer (GAT), Tono-Pen, and noncontact tonometer (NCT) in the measurement of intraocular pressure (IOP) in pseudophakic children.

Methods: The medical records of nonglaucomatous pseudophakic children between 2009 and 2019 were retrospectively analyzed. A total of 46 eyes of 23 patients operated for bilateral pediatric cataract were included in the study. The patients' mean age was 13.4 ± 4.1 years. Central corneal thickness (CCT) and IOP values measured with the GAT, Tono-Pen, and NCT were recorded. Agreement between the tonometers was evaluated by intraclass correlation coefficients (ICC) and the Bland-Altman method.

Results: The mean IOP of the 46 eyes included in the study was measured as 13.7 ± 2.3 mm Hg with the GAT, 16.0 ± 2.3 mm Hg with NCT, and 16.5 ± 2.3 mm Hg with the Tono-Pen (p < 0.001). There is no statistical difference between NCT and Tono-Pen measurements, while GAT measurements were significantly lower than those of the NCT and Tono-pen. ICC values showed fair agreement between NCT and Tono-Pen (ICC = 0.720), whereas there was poor agreement between GAT and NCT (ICC = 0.501) and Tono-pen (ICC = 0.314).

Conclusions: With all devices included in the study, thicker corneas were associated with higher IOP measurements. Although there was moderate agreement between the NCT and Tono-Pen, there was a statistically significant difference in the IOP values provided by the three devices. Our results suggest these devices should not be used interchangeably.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10792-024-03210-wDOI Listing

Publication Analysis

Top Keywords

intraocular pressure
8
applanation tonometer
8
noncontact tonometer
8
nonglaucomatous pseudophakic
8
pseudophakic children
8
gat tono-pen
8
included study
8
comparison intraocular
4
pressure measurements
4
tono-pen
4

Similar Publications

Neurodegeneration in glaucoma patients is clinically identified through longitudinal assessment of structure-function changes, including intraocular pressure, cup-to-disc ratios from fundus images, and optical coherence tomography imaging of the retinal nerve fiber layer. Use of human post-mortem ocular tissue for basic research is rising in the glaucoma field, yet there are challenges in assessing disease stage and severity, since tissue donations with informed consent are often unaccompanied by detailed pre-mortem clinical information. Further, the interpretation of disease severity based solely on anatomical and morphological assessments by histology can be affected by differences in death-to-preservation time and tissue processing.

View Article and Find Full Text PDF

Purpose: To investigate retinal microvascular changes in ischemic stroke patients using optical coherence tomography angiography (OCT-A) and assess these alterations based on stroke etiology.

Methods: Case-control study conducted at Montpellier University Hospital from May 2021 to March 2022 (IRB: 202000607). Retinal vascular features were compared between strokes patients and age- and sex- matched controls.

View Article and Find Full Text PDF

Removal of Subretinal Gas After Pneumatic Retinopexy Under Air Using External Manipulation.

Retin Cases Brief Rep

January 2025

The Retina Service of Wills Eye Hospital, Wills Eye Physicians-Mid Atlantic Retina, Thomas Jefferson University, Philadelphia, PA.

Purpose: To illustrate a technique for the removal of subretinal gas via pars plana vitrectomy (PPV) with air-fluid exchange and simultaneous manipulation with scleral depression.

Methods: PPV to remove subretinal gas causing persistent macula-off retinal detachment was performed in one eye, and the results were evaluated in this case report. Ports were carefully placed to avoid puncturing the retina, which was significantly displaced anteriorly past the ora serrata due to the buoyancy of the subretinal gas with the patient in a supine position.

View Article and Find Full Text PDF

Ischemia-reperfusion injuries are known to cause a range of retinal pathologies, including diabetic retinopathy, glaucoma, retinal vascular occlusions, and other vaso-occlusive conditions. This manuscript presents a method for inducing ischemia-reperfusion injury in a mouse model. The method utilized anterior chamber cannulation attached to a saline reservoir, generating hydrostatic pressure to raise the intraocular pressure to 90-100 mmHg.

View Article and Find Full Text PDF

Compartmental analysis of retinal vascular parameters and thickness in myopic eyes using SS-OCTA.

Front Med (Lausanne)

December 2024

Chongqing Key Laboratory of Prevention and Treatment on Major Blinding Diseases, Chongqing Eye Institute, Chongqing Branch (Municipality Division) of National Clinical Research Center for Ocular Diseases, The First Affiliated Hospital of Chongqing Medical University, Chongqing, China.

Background: This study aimed to comprehensively explore the thickness and topographic distributions of retinal vessel alterations of different myopic eyes by using swept-source OCT angiography (SS-OCTA).

Methods: One hundred myopes were included in this observational cross-sectional study. All participants underwent a series of ocular examinations of biometrical parameters, including spherical equivalent refraction (SER), axial length (AL), intraocular pressure (IOP), curvature radius (CR), and others.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!