Purpose: This study aimed to compare the image quality of chest computed tomography (CT) scans for COVID-19 pneumonia using forward-projected model-based iterative reconstruction solution-LUNG (FIRST-LUNG) with filtered back projection (FBP) and hybrid iterative reconstruction (HIR).

Method: The CT images of 44 inpatients diagnosed with COVID-19 pneumonia between December 2022 and June 2023 were retrospectively analyzed. The CT images were reconstructed using FBP, HIR, and FIRST-LUNG-MILD/STANDARD/STRONG. The CT values and noise of the lumen of the main trachea and erector spine muscle were measured for each group. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated. Subjective evaluations included overall image quality, noise, streak artifact, visualization of normal lung structures, and abnormal CT features. One-way analysis of variance was used to compare the objective and subjective indicators among the five groups. The task-based transfer function was derived for three distinct contrasts representing anatomical structures, lower-contrast lesion, and higher-contrast lesion.

Results: The results of the study demonstrated significant differences in image noise, SNR, and CNR among the five groups ( P < 0.001). The FBP images exhibited the highest levels of noise and the lowest SNR and CNR among the five groups ( P < 0.001). When compared to the FBP and HIR groups, the noise was lower in the FIRST-LUNG-MILD/STANDARD/STRONG group, while the SNR and CNR were higher ( P < 0.001). The subjective overall image quality score of FIRST-LUNG-MILD/STANDARD was significantly better than FBP and FIRST-LUNG-STRONG ( P < 0.001). FIRST-LUNG-MILD was superior to FBP, HIR, FIRST-LUNG-STANDARD, and FIRST-LUNG-STRONG in visualizing proximal and peripheral bronchovascular and subpleural vessels ( P < 0.05). Additionally, FIRST-LUNG-MILD achieved the best scores in evaluating abnormal lung structure ( P < 0.001). The overall interobserver agreement was substantial (intraclass correlation coefficient = 0.891). The task-based transfer function 50% values of FIRST reconstructions are consistently higher compared to FBP and HIR.

Conclusions: The FIRST-LUNG-MILD/STANDARD algorithm can enhance the image quality of chest CT in patients with COVID-19 pneumonia, while preserving important details of the lesions, better than the FBP and HIR algorithms. After evaluating various COVID-19 pneumonia lesions and considering the improvement in image quality, we recommend using the FIRST-LUNG-MILD reconstruction for diagnosing COVID-19 pneumonia.

Download full-text PDF

Source
http://dx.doi.org/10.1097/RCT.0000000000001635DOI Listing

Publication Analysis

Top Keywords

image quality
24
covid-19 pneumonia
24
fbp hir
16
iterative reconstruction
12
quality chest
12
snr cnr
12
model-based iterative
8
chest computed
8
computed tomography
8
fbp
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!