Total and simulated keratometry measurements using IOLMaster 700 and Pentacam AXL after small incision lenticule extraction.

Int Ophthalmol

Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, No.1 Dongjiaomin Ln, Dongcheng District, Beijing, 100730, China.

Published: June 2024

Purpose: Calculating the intraocular lens (IOL) in patients after corneal refractive surgery presents a challenge. Because an overestimation of corneal power in cases undergone this surgery leading to a subsequent under-correction of IOL power. However, recent advancements in technology have eliable measurement of total corneal power. The aim of this research was to assess the agreement in simulated keratometry (SimK) and total keratometry (TK) values between IOLMaster 700 and Pentacam AXL.

Methods: The study involved 99 patients (99 eyes) undergone small incision lenticule extraction (SMILE) surgery. Each patient underwent scans using IOL Master 700 and Pentacam AXL. The following parameters were recorded: SimK1, SimK2, Total K1 (TK1), and Total K2 (TK2) for IOLMaster 700; and SimK1, SimK2, True Net Power (TNP) K1, TNPK2, Total Corneal Refractive Power (TCRP) K1, and TCRP K2 for Pentacam AXL. Agreement between the two devices was evaluated using Bland-Altman plot, while paired t-test was utilized to compare any differences in the same parameter by both instruments.

Results: The results revealed a strong correlation between the two devices.Noticeable comparability was identified for all SimK variables. However, there were noticeable differences in TK measurements as well as TK1-TNPK1, TK2-TNP K2, TK1-TCRP K1, and TK2-TCRP K2 parameters when comparing the two devices. The IOLMaster 700 consistently measured steeper values than the Pentacam AXL, with significant and clinically relevant differences of 1.34, 1.37, 0.87, and 0.95 diopters, respectively.

Conclusion: While there was a noticeable correlation between the IOLMaster 700 and Pentacam AXL in SimK measurements, a marked difference was noted in TK values. The two devices cannot be used interchangeably when quantifying TK values.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10792-024-03191-wDOI Listing

Publication Analysis

Top Keywords

iolmaster 700
20
pentacam axl
20
700 pentacam
16
simulated keratometry
8
small incision
8
incision lenticule
8
lenticule extraction
8
corneal refractive
8
corneal power
8
total corneal
8

Similar Publications

Changes in Axial Length in Patients of Acute Retinal Necrosis Undergoing Vitrectomy.

Ocul Immunol Inflamm

January 2025

Eye Institute and Department of Ophthalmology, Eye and ENT Hospital, Fudan University, Shanghai, China.

Purpose: To observe the changes in the axial length (AL) in patients of acute retinal necrosis (ARN) undergoing vitrectomy and investigate the correlated factors.

Methods: Retrospective case series. Patients diagnosed as ARN undergoing vitrectomy with silicone oil (SO) tamponade, and with attached retina more than one year after silicone oil removal (SOR) were included.

View Article and Find Full Text PDF

Purpose: To apply a new method, the Taylor Diagram, and a new concept, the centered root mean square error (cRMSE), in evaluating the performance of IOL formulas.

Methods: The preoperative biometrics were measured with the IOLMaster 700 (Carl Zeiss Meditec) and the postoperative spherical equivalent refraction (SER) was calculated in 888 anonymous patients. The Taylor Diagram was applied to visualize the centered root mean square error (cRMSE) and the correlation coefficient between the predictions and the observations (R).

View Article and Find Full Text PDF

Purpose: To evaluate the accuracy of toric intraocular lens (IOL) axis prediction between two preoperative measurement devices: the optical biometry (IOLMaster 500 or 700) and the dual Scheimpflug topography (Galilei G4).

Methods: Medical records of 64 eyes from 44 patients who underwent phacoemulsification and posterior chamber toric IOL (Zeiss AT TORBI 709M) implantation between July 2017 and January 2022 were reviewed. All patients underwent preoperative evaluation by optical biometry (IOLMaster 500 or IOLMaster 700) and Galilei G4.

View Article and Find Full Text PDF

Purpose: To investigate different measures for corneal astigmatism in the context of reconstructed corneal astigmatism (recCP) as required to correct the pseudophakic eye, and to derive prediction models to map measured corneal astigmatism to recCP.

Methods: Retrospective single centre study of 509 eyes of 509 cataract patients with monofocal (MX60P) IOL. Corneal power measured with the IOLMaster 700 keratometry (IOLMK), and Galilei G4 keratometry (GK), total corneal power (TCP2), and Alpin's integrated front (CorT) and total corneal power (CorTTP).

View Article and Find Full Text PDF

Purpose: This study aims to compare and evaluate the correlation and agreement among three distinct anterior segment analysis systems in measuring anterior chamber depth (ACD) in healthy myopic individuals.

Methods: Myopic patients who underwent preoperative assessments at our hospital's refractive surgery department from January 2021 to December 2023 were included in this investigation. ACD was measured using three anterior segment analysis systems: Pentacam, Sirius, and IOLMaster 700.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!