Background: Numerous immunoassays have been commercialized to determine pancreatic elastase (PE) in feces in screening for exocrine pancreatic insufficiency (EPI), but how the different assays compare to one another is controversial, especially in the context that all methods use the same cut-off values for interpreting the results obtained on the presence or absence of EPI or the degree of insufficiency if it is present. Our aim was to analytically verify a new method for determining PE, compare the results with a previous method, and verify the declared cut-off values for interpretation of the results.

Methods: PE in the stool was assayed using a previous monoclonal enzyme-linked immunosorbent assay ("ScheBo ELISA") and a new polyclonal particle-enhanced turbidimetric immunoassay ("Bühlmann PETIA"). The direct method comparison of two immunoassays was performed in 40 samples. Clinical comparisons were conducted against each other for the binary determination of "abnormal/normal" elastase levels and the three-way determination of "severe/moderate/no" EPI in 56 samples. The indirect comparison method used external quality assessment (EQA) data to compare the monoclonal and polyclonal immunoassays for PE, and additionally compare the monoclonal ScheBo ELISA to a monoclonal chemiluminescence immunoassay ("DiaSorin CLIA").

Results: Precision in the series and intra-laboratory precision for Bühlmann PETIA met the manufacturer's specifications for the concentration range of limit/lower values and the range of normal values. The Bühlmann PETIA immunoassay on different analytical platforms yielded comparable results and nearly perfect agreement in the case of three-way classification (kappa = 0.89 with 95%CI from 0.79 to 1.00. ScheBo ELISA tends to generate higher values of pancreatic elastase than the Bühlmann PETIA; agreement between the methods was moderate in the case of binary classification (kappa = 0.43; 95% CI 0.25 to 0.62), and substantial in the case of three-way classification (kappa = 0.62; 95% CI 0.50 to 0.75). EQA data analysis showed a statistically significant difference between ScheBo ELISA and Bühlmann PETIA peer groups ( = 0.031), as well as the DiaSorin CLIA and ScheBo ELISA peer groups ( = 0.010).

Conclusion: The ScheBo ELISA and Bühlmann PETIA do not appear to be commutable in the analytical and clinical context. Our data address a discordance between different mono- and polyclonal immunoassays for pancreatic elastase and the potential of misclassification using its universal cut-off values in screening suspected patients for exocrine pancreatic insufficiency.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11171928PMC
http://dx.doi.org/10.3390/diagnostics14111166DOI Listing

Publication Analysis

Top Keywords

schebo elisa
20
bühlmann petia
20
pancreatic elastase
16
polyclonal immunoassays
12
cut-off values
12
classification kappa
12
monoclonal polyclonal
8
exocrine pancreatic
8
pancreatic insufficiency
8
eqa data
8

Similar Publications

Background: Numerous immunoassays have been commercialized to determine pancreatic elastase (PE) in feces in screening for exocrine pancreatic insufficiency (EPI), but how the different assays compare to one another is controversial, especially in the context that all methods use the same cut-off values for interpreting the results obtained on the presence or absence of EPI or the degree of insufficiency if it is present. Our aim was to analytically verify a new method for determining PE, compare the results with a previous method, and verify the declared cut-off values for interpretation of the results.

Methods: PE in the stool was assayed using a previous monoclonal enzyme-linked immunosorbent assay ("ScheBo ELISA") and a new polyclonal particle-enhanced turbidimetric immunoassay ("Bühlmann PETIA").

View Article and Find Full Text PDF

The clinical and laboratory impact of reporting faecal elastase-1 in watery stool samples.

Ann Clin Biochem

January 2023

Department of Clinical Biochemistry, 355202York Teaching Hospital, York, YO31 8HE, UK.

Background: Low faecal elastase-1 (FE-1) results are suggestive of pancreatic insufficiency, but watery diarrhoea may lead to falsely low results.

Methods: FE-1 results reported on watery samples over a three-year period were reviewed. Results in watery samples were compared to those from a formed sample.

View Article and Find Full Text PDF

Background: Fecal pancreatic elastase 1 (FPE1) is an established screening test for pancreatic exocrine insufficiency (PEI), a condition that is underdiagnosed and if not treated may cause significant morbidity. The aim of this study was to compare a new FPE1 machine based CLIA kit to an ELISA assay which is considered the de facto gold standard in our laboratory for FPE1 measurement.

Methods: Levels of FPE1 from the 227 stool samples were analyzed by the ScheBo ELISA kit and the CLIA Liaison XL system simultaneously with the same cutoff values for both assays.

View Article and Find Full Text PDF

Objectives: Fecal pancreatic elastase (PE) assays are screening tests for exocrine pancreatic insufficiency (EPI). We analytically evaluated a new PE assay and retrospectively analyzed data from an academic hospital and reference laboratory to understand the clinical utility.

Methods: Forty stool samples with different PE concentrations were tested on the ScheBo enzyme-linked immunosorbent assay (ELISA) versus DiaSorin LIAISON immunoassay; a simple-to-use extraction device was assessed.

View Article and Find Full Text PDF

Background: Tumor pyruvate kinase M2 isoform (tM2-PK), which is an isoform of PK-glycolytic enzyme and appears on the surface of cancerous proliferating cells, has been used as a diagnostic biomarker for colorectal cancer (CRC). The aim of this study was to evaluate the tM2-PK measurement test for the diagnosis of CRCs and adenomatous polyps in plasma and stool samples in an Iranian population.

Methods: In this prospective study, a total of 226 stool and 178 plasma samples were received from patients referred to colonoscopy units.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!