Endotracheal intubation (ETI) is a cornerstone of airway management. The gold standard device for ETI is still the direct laryngoscope (DL). However, video laryngoscopes (VLs) are now also widely available and have several proven advantages. The VL technique has been included in the major airway management guidelines. During the COVID-19 pandemic, supply chain disruption has raised demand for 3D-printed medical equipment, including 3D-printed VLs. However, studies on performance are only sparsely available; thus, we aimed to compare 3D-printed VLs to the DL and other VLs made with conventional manufacturing technology. Forty-eight medical students were recruited to serve as novice users. Following brief, standardized training, students executed ETI with the DL, the King Vision (KV), the VividTrac (VT), the AirAngel Blade (AAB), and a custom-made 3D-printed VL (3DVL) on the Laerdal airway management trainer in normal and difficult airway scenarios. We evaluated the time to and proportion of successful intubation, the best view of the glottis, esophageal intubation, dental trauma, and user satisfaction. The KV and VT are proved to be superior ( < 0.05) to the DL in both scenarios. The 3DVL's performance was similar ( > 0.05) or significantly better than that of the DL and mainly non-inferior ( > 0.05) compared to the KV and VT in both scenarios. Regardless of the scenario, the AAB proved to be inferior ( < 0.05) even to the DL in the majority of the variables. The differences between the devices were more pronounced in the difficult airway scenario. The user satisfaction scores were in concordance with the aforementioned performance of the scopes. Based upon our results, we cannot recommend the AAB over the DL, KV, or VT. However, as the 3DVL showed, 3D printing indeed can provide useful or even superior VLs, but prior to clinical use, meticulous evaluation might be recommended.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11173105PMC
http://dx.doi.org/10.3390/jcm13113213DOI Listing

Publication Analysis

Top Keywords

airway management
12
king vision
8
vision vividtrac
8
vividtrac airangel
8
airangel blade
8
custom-made 3d-printed
8
3d-printed vls
8
difficult airway
8
user satisfaction
8
3d-printed
5

Similar Publications

Background: Prolonged dependence on mechanical ventilation is a common occurrence in clinical ICU patients and presents significant challenges for patient care and resource allocation. Predicting prolonged dependence on mechanical ventilation is crucial for improving patient outcomes, preventing ventilator-associated complications, and guiding targeted clinical interventions. However, specific tools for predicting prolonged mechanical ventilation among ICU patients, particularly those with critical orthopaedic trauma, are currently lacking.

View Article and Find Full Text PDF

Comparative efficacy and safety of pulmonary surfactant delivery strategies in neonatal RDS: a network meta-analysis.

BMC Pulm Med

December 2024

Department of Neonatology, Children's Hospital of Chongqing Medical University, Chongqing, 400,014, China.

Purpose: To compare five pulmonary surfactant (PS) administration strategies for neonates with respiratory distress syndrome (RDS), including intubation-surfactant-extubation (InSurE), thin catheter administration, laryngeal mask airway (LMA), surfactant nebulization (SN), and usual care, with a particular emphasis on the comparison of the LMA and SN with other strategies.

Methods: We conducted a systematic search of MEDLINE, EMBASE, PUBMED, and Cochrane CENTRAL databases up to November 2023. Two authors independently conducted data extraction, and assessed bias using the Cochrane Risk of Bias Tool.

View Article and Find Full Text PDF

Background: We previously published the protocol and statistical analysis plan for a randomized controlled trial of Proportional Assist Ventilation for Minimizing the Duration of Mechanical Ventilation: the PROMIZING study in Trials ( https://doi.org/10.1186/s13063-023-07163-w ).

View Article and Find Full Text PDF

Background: Considerable morbidity is attributable to inappropriate tracheal cuff pressure. An earlier study undertaken in our hospital revealed that a normal cuff pressure of 20-30 cm HO was achieved in only 6% of intubated patients using subjective estimation methods.

Objective: To determine whether a training intervention could improve the accuracy of the subjective estimation method in our tracheal cuff monitoring.

View Article and Find Full Text PDF

Arteriovenous malformations (AVMs) in the head and neck present significant challenges due to airway management complexities and hemorrhage risks. This case report describes a 15-year-old female with a congenital facial AVM causing dyspnea and obstructive symptoms. The patient required angioembolization of the AVM, but many hospitals deferred the procedure due to the anticipated difficult airway and severe bleeding risks.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!