A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Improving diagnosis of early complications (<1 week) through continuous vital sign monitoring following oncological gastrointestinal surgical procedures. | LitMetric

Background: Patients undergoing major oncological abdominal surgery are prone to postoperative complications, making early recognition crucial. Clinical deterioration is often preceded by changes in vital signs, which are typically measured thrice a day by a nurse. However, intermittent measurements may delay recognizing clinical deterioration. Continuous vital parameter monitoring may lead to earlier recognition and management of complications and reduce nursing workload.

Objective: To compare vital parameter measurements between ward nurses and a wireless continuous monitoring system (Sensium® wireless patch) and assess whether this patch can detect clinical deterioration earlier in patients with complications in the first postoperative week.

Methods: Vital parameters (heart rate, respiratory rate, and temperature) were collected in patients undergoing an oncological resection of the liver, colorectal, or pancreas. Sensium® patch measurements were compared to nurses' measurements to assess the percentages of discordant measurements. In patients with complications in the first postoperative week, time discrepancies between nurses and Sensium® patch measurements were identified in cases of clinical deterioration (respiratory rate ≥15/min, heart rate ≥100/min, and temperature ≥38°C).

Results: Among 227 patients, 22% of the patients experienced complications. Nurse and Sensium® measurements were discrepant in 586/2272 measurements (26%). In 506/586 discrepancies (86%), this was due to the respiratory rate (difference ≥4/min). Compared to nurses, the Sensium® patch detected an elevated respiratory rate 14 h earlier and heart rate 2 h earlier within complications in the first postoperative week. For temperature, no difference was observed.

Conclusion: Continuous monitoring with the Sensium® wireless patch holds promise for earlier recognition of complications in patients who underwent major oncological abdominal surgery.

Download full-text PDF

Source
http://dx.doi.org/10.1002/wjs.12248DOI Listing

Publication Analysis

Top Keywords

improving diagnosis
4
diagnosis early
4
early complications
4
improving
1
early
1
complications
1

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!