Background: Enteral feeding pump systems deliver decreased amounts of macronutrients in human milk to neonates. This study determined the macronutrient loss associated with a bottle-feeding pump system and the effect of manually mixing the human milk during extended feeds.

Methods: Macronutrient content from samples of donor human milk was analyzed after simulated extended feeds with a bottle-feeding pump system, using a human milk analyzer. Simulations were repeated using manual mixing of the bottle every 30 min during feeding. The percentage of the baseline was calculated, and one-sample t tests and analysis of variance compared the effect of manual mixing and the duration of feeding on macronutrient delivery.

Results: The delivery of fat and energy was lower over time, but manual mixing considerably improved retention. The length of feeding impacted fat delivery, with less fat delivered over time (P < 0.001). Manually mixing significantly increased fat delivery (P < 0.001). Similar results were found for energy, with a significant reduction in energy delivery over time (P < 0.001) and significantly more energy delivered with mixing (P < 0.001). Mixing and the duration of feeding had minimal effect on protein or carbohydrate delivery.

Conclusions: Bottle-feeding pump systems are associated with a significant reduction in the delivery of fat and energy of donor human milk. The manual mixing of donor human milk during prolonged feeds is a simple way to improve fat and energy delivery to the neonate.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11293984PMC
http://dx.doi.org/10.1002/jpen.2659DOI Listing

Publication Analysis

Top Keywords

human milk
20
bottle-feeding pump
12
manual mixing
12
fat energy
8
pump system
8
delivery fat
8
human
5
milk
5
simple step
4
step improve
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!