Artifactually increased serum bicarbonate in a cat with rhabdomyolysis.

Vet Clin Pathol

Department of Veterinary Medicine, Queen's Veterinary School Hospital, University of Cambridge, Cambridge, UK.

Published: September 2024

A 3-year-old male neutered domestic shorthair cat presented with lethargy, hyporexia, and pyrexia of unknown origin. Biochemical analysis using a Beckman Coulter AU480 demonstrated marked increases in creatine kinase and aspartate aminotransferase, indicative of severe muscle injury, with concurrent presumptive myoglobinuria on urinalysis. A marked, non-physiologic increase in measured bicarbonate and resultant negative anion gap was documented; however, calculated bicarbonate obtained via a point-of-care blood gas analyzer was within normal limits. Laboratory error due to interference by lactate dehydrogenase was suspected and supported by the results of subsequent biochemical testing. Artifactual increases in bicarbonate have been documented in cases of rhabdomyolysis in horses, cows, and a bird. However, to the best of the authors' knowledge, this is the first report to demonstrate this spurious change in a cat.

Download full-text PDF

Source
http://dx.doi.org/10.1111/vcp.13371DOI Listing

Publication Analysis

Top Keywords

artifactually increased
4
increased serum
4
bicarbonate
4
serum bicarbonate
4
bicarbonate cat
4
cat rhabdomyolysis
4
rhabdomyolysis 3-year-old
4
3-year-old male
4
male neutered
4
neutered domestic
4

Similar Publications

Artifacts at Cardiac MRI: Imaging Appearances and Solutions.

Radiographics

January 2025

From the Department of Radiology, Cardiovascular Imaging, Mayo Clinic, 200 1st St SW, Rochester, MN 559905 (P.S.R., P.A.A.); Department of Radiology, Division of Cardiothoracic Imaging, Jefferson University Hospitals, Philadelphia, Pa (B.S.); Department of Radiology, Baylor Health System, Dallas, Tex (P.R.); Department of Diagnostic Radiology, School of Clinical Medicine, Li Ka Shing Faculty of Medicine, University of Hong Kong, Hong Kong SAR (M.Y.N.); and Department of Diagnostic Radiology, Cleveland Clinic, Cleveland, Ohio (M.A.B.).

Cardiac MRI (CMR) is an important imaging modality in the evaluation of cardiovascular diseases. CMR image acquisition is technically challenging, which in some circumstances is associated with artifacts, both general as well as sequence specific. Recognizing imaging artifacts, understanding their causes, and applying effective approaches for artifact mitigation are critical for successful CMR.

View Article and Find Full Text PDF

Background: Minimizing air aspiration by carefully filling blood gas syringes is crucial to prevent air contamination from causing undesirable variations in gasses and other molecules. While some previous studies investigated this aspect, these are now outdated and only analyzed a limited number of blood gas parameters. Thus, we investigated the effects air contamination in the syringe using a modern blood gas analyzer.

View Article and Find Full Text PDF

Bearded dragons () are a common reptile species kept under human care and suffer from a wide range of diseases for which plasma biochemistry is used as a first-line diagnostic test. There is limited information available regarding tissue enzyme activities and origin that could assist in interpreting the bearded dragon plasma biochemistry enzymology profile. The aim of this study was to characterize the tissue activities of seven enzymes routinely used in the reptile biochemistry panel: alkaline phosphatase (ALP), alanine aminotransferase (ALT), aspartate aminotransferase (AST), gamma-glutamyl transferase (GGT), glutamate dehydrogenase (GLDH), lactate dehydrogenase (LDH), and creatine kinase (CK) in 12 adult inland bearded dragons in 13 tissues, plasma, and red blood cells.

View Article and Find Full Text PDF

Background: The aetiology of increased mean platelet volume in dogs is currently uncertain. Our aim was to investigate the aetiology of increased mean platelet volume in dogs with and without thrombocytopenia.

Methods: The database of a teaching hospital was retrospectively searched for dogs with increased mean platelet volume (>14.

View Article and Find Full Text PDF
Article Synopsis
  • The study aimed to explore how contaminating venous blood samples with lipid solutions affects measurements from blood gas analyzers.
  • Blood samples were collected from 17 healthcare workers with varying percentages (0%, 5%, and 10%) of lipid solution, and analyzed quickly for blood gas parameters and triglyceride levels.
  • Results showed that higher lipid contamination significantly altered various analytes, indicating that contamination can lead to clinically relevant inaccuracies in blood gas analysis, suggesting the need for improved blood gas analyzers that include serum indices.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!