Setting the Separation Factor α for Ketone Monomers and Dimers by the Use of Different Drift Gases.

J Am Soc Mass Spectrom

Miniaturisation, Leibniz-Institut für Analytische Wissenschaften ISAS e.V., 44139 Dortmund, Germany.

Published: July 2024

This study investigates the influence of different drift gases on ion mobility in ion mobility spectrometry (IMS) using ketones as model substances within a custom-built drift tube spectrometer. Different binary mixtures of nitrogen, helium, and argon were used as drift gases to investigate the influence of mobility on the monomers and dimers of the different ketones. Experimental results reveal shifts in ion drift times and separation factors (α) with varying gas compositions, in accordance with Blanc's Law. Furthermore, the study underscores the device-independent nature of α and the device-dependent resolution, emphasizing the importance of comparative analyses. Employing 2-hexanone and 2-decanone in the same sample but with different drift gases is used to show the impact of different drift gases. By changing the drift gas composition, total alignment of drift times and therefore no possible resolution or baseline resolution could be achieved. Through different experiments and analyses, this research provides insights into the interactions between gas composition and ion mobility, offering implications for diverse analytical applications from environmental monitoring to chemical detection.

Download full-text PDF

Source
http://dx.doi.org/10.1021/jasms.4c00215DOI Listing

Publication Analysis

Top Keywords

drift gases
20
ion mobility
12
drift
9
monomers dimers
8
drift times
8
gas composition
8
gases
5
setting separation
4
separation factor
4
factor ketone
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!