Early wound intervention and closure is critical for reducing infection and improving aesthetic and functional outcomes for patients with acute burn wounds and nonthermal full-thickness skin defects. Treatment of partial-thickness burns or full-thickness injuries with autologous skin cell suspension (ASCS) achieves robust wound closure while limiting the amount of donor skin compared with standard autografting. A Next Generation Autologous Cell Harvesting Device (NG-ACHD) was developed to standardize the preparation process for ASCS to ensure biological attributes are obtained known to correlate with well-established safety and performance data. This study compared ASCS prepared using the NG-ACHD and ACHD following the manufacturer's guidance, evaluating cellular yields, viability, apoptotic activity, aggregates, phenotypes and functional capacity. Non-inferiority was established for all biological attributes tested and comparable healing trajectories were demonstrated using an in vitro skin regeneration model. In addition to standardization, the NG-ACHD also provides workflow efficiencies with the potential to decrease training requirements and increase the ease of incorporation and utilization of ASCS in clinical practice.