Purpose: This study leverages externally generated Pilot Tone (PT) signals to perform motion-corrected brain MRI for sequences with arbitrary k-space sampling and image contrast.

Theory And Methods: PT signals are promising external motion sensors due to their cost-effectiveness, easy workflow, and consistent performance across contrasts and sampling patterns. However, they lack robust calibration pipelines. This work calibrates PT signal to rigid motion parameters acquired during short blocks (˜4 s) of motion calibration (MC) acquisitions, which are short enough to unobstructively fit between acquisitions. MC acquisitions leverage self-navigated trajectories that enable state-of-the-art motion estimation methods for efficient calibration. To capture the range of patient motion occurring throughout the examination, distributed motion calibration (DMC) uses data acquired from MC scans distributed across the entire examination. After calibration, PT is used to retrospectively motion-correct sequences with arbitrary k-space sampling and image contrast. Additionally, a data-driven calibration refinement is proposed to tailor calibration models to individual acquisitions. In vivo experiments involving 12 healthy volunteers tested the DMC protocol's ability to robustly correct subject motion.

Results: The proposed calibration pipeline produces pose parameters consistent with reference values, even when distributing only six of these approximately 4-s MC blocks, resulting in a total acquisition time of 22 s. In vivo motion experiments reveal significant ( ) improved motion correction with increased signal to residual ratio for both MPRAGE and SPACE sequences with standard k-space acquisition, especially when motion is large. Additionally, results highlight the benefits of using a distributed calibration approach.

Conclusions: This study presents a framework for performing motion-corrected brain MRI in sequences with arbitrary k-space encoding and contrast, using externally generated PT signals. The DMC protocol is introduced, promoting observation of patient motion occurring throughout the examination and providing a calibration pipeline suitable for clinical deployment. The method's application is demonstrated in standard volumetric MPRAGE and SPACE sequences.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mrm.30161DOI Listing

Publication Analysis

Top Keywords

sequences arbitrary
12
arbitrary k-space
12
motion
10
calibration
10
pilot tone
8
tone signals
8
externally generated
8
motion-corrected brain
8
brain mri
8
mri sequences
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!