Endoscopic angle-resolved light scattering methods have been developed for early cancer detection but they typically require multi-element coherent fiber optic bundles to recover scattering distributions from tissues. Recent work has focused on using a single multimode fiber (MMF) to measure angle resolved scattering but this approach has practical limitations to overcome before clinical translation. Here we address these limitations by proposing an MMF-based endoscope capable of measuring angular scattering patterns suitable for determining structure. Significantly, this approach implements a spectrally resolved detection scheme to reduce speckle and leverages the azimuthal symmetry of the angular scattering patterns to enable measurements that are robust to fiber bending. This results in a unique method that does not require matrix inversion or machine learning to measure a transmitted scattering distribution. The MMF utilized here is 1000 mm in length with a 200 µm core and is demonstrated to recover angular scattering distributions even with bending displacements of up to 30 cm. This advance has a significant impact on the clinical translation of biomedical endoscopic diagnostic techniques that use angular scattering to determine the size of cell nuclei to detect early cancer.