A spectrometer built using an external cavity pulsed quantum cascade laser is described. The spectrometer has a tuning range from 10 - 13 µm (1,000 - 769 cm) and is designed to target volatile organic compounds (VOCs) which often exhibit water-free molecular absorption within the region. The spectrometer utilizes a hollow silica waveguide gas cell which has an internal volume of a few millilitres, a fast response time (∼1 s), and is advantageous when only low sample volumes, similar to the cell volume, are available. Propane is used as a test gas because it is easy to handle, and its spectral profile is comparable to VOCs of interest. Its absorption in the region is primarily within the ν band which spans from 10.55 - 11.16 µm (948 - 896 cm). Spectral measurements at a range of concentrations show good linearity and an Allan deviation of absorbance values recorded over a 100-minute period indicates a minimum detectable absorbance of 3.5×10 at an integration time of 75 s.