Endovaginal coil for pelvic high-resolution magnetic resonance imaging of cervical cancer: a preliminary parameter optimization study.

Quant Imaging Med Surg

State Key Laboratory of Ultrasound in Medicine and Engineering, College of Biomedical Engineering, Chongqing Medical University, Chongqing, China.

Published: June 2024

Background: The diagnosis of early-stage cervical cancer through conventional magnetic resonance imaging (MRI) remains challenging, highlighting a greater need for pelvic high-resolution MRI (HR MRI). This study used our research team's endovaginal coil imaging to optimize scanning parameters and aimed to achieve HR MRI of the pelvis and determine its clinical value.

Methods: Fifty participants were recruited prospectively for this cross-sectional study conducted at the First Affiliated Hospital of Chongqing Medical University from January 2023 to November 2023. Initially, 10 volunteers requiring pelvic imaging diagnosis underwent pelvic MRI with the endovaginal coil combined with a conventional external array coil to test and optimize the scanning parameters. Subsequently, 40 patients who were highly suspected or diagnosed with cervical cancer were randomly assigned to undergo an initial pelvic scan with an external array coil with subsequent examinations of both the conventional coil and the endovaginal coil. Two experienced radiologists performed quantitative analyses, measuring signals and calculating the signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and contrast (C). They also conducted qualitative analyses, evaluating imaging artifacts, anatomical structures, and overall image quality. The paired sample -test and Wilcoxon rank-sum test were conducted to compare the statistical differences between the two sets of images, while the intraclass correlation coefficient (ICC) and Kappa consistency tests were used to assess the measurement and scoring consistency between the two radiologists.

Results: The optimized endovaginal images had higher mean SNR, CNR, and C values (18.62±7.85, 16.04±7.72, and 0.73±0.11, respectively) compared to the conventional images (6.77±2.36, 4.47±2.05, and 0.47±0.12, respectively). Additionally, the ratings for imaging artifacts, anatomical structures, and overall quality of the endovaginal images were all 4 [interquartile range (IQR) 4, 4]; meanwhile, the conventional images scored lower with ratings of 4 (IQR 3, 4), 3 (IQR 3, 3), and 3 (IQR 3, 3) for SNR, CNR, and C, respectively. All analysis results underwent paired-sample -tests or Wilcoxon rank-sum tests between the two groups, yielding a P value <0.001. The optimized endovaginal images also showed improved resolution with a reconstructed voxel size of 0.11 mm, and HR MRI was successfully achieved. The ICC values for the measurements were 0.914, 0.947, and 0.912, respectively, and for the ratings, the measurement was 0.923, indicating excellent consistency between the two physicians (ICC/Kappa value between 0.85 and 1.00).

Conclusions: Endovaginal technology, which provides precise clinical information for the diagnosis of cervical cancer, provides straightforward operation and exceptional imaging quality, making it highly suitable for expanded clinical use.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11151224PMC
http://dx.doi.org/10.21037/qims-23-1718DOI Listing

Publication Analysis

Top Keywords

endovaginal coil
16
cervical cancer
12
pelvic high-resolution
8
magnetic resonance
8
resonance imaging
8
optimize scanning
8
scanning parameters
8
external array
8
array coil
8
imaging artifacts
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!