Background: Electrographic flow (EGF) mapping enables full spatiotemporal reconstruction of organized wavefront propagation to identify extrapulmonary vein sources of atrial fibrillation (AF).
Objectives: FLOW-AF (A Randomized Controlled Study to Evaluate the Reliability of the Ablacon Electrographic FLOW [EGF] Algorithm Technology [Ablamap Software] to Identify AF Sources and Guide Ablation Therapy in Patients With Persistent Atrial Fibrillation) was multicenter, randomized controlled study of EGF mapping to: 1) stratify a nonparoxysmal AF population undergoing redo ablation; 2) guide ablation of these extrapulmonary vein AF sources; and 3) improve AF recurrence outcomes.
Methods: FLOW-AF enrolled persistent atrial fibrillation (PerAF)/long-standing PerAF patients undergoing redo ablation at 4 centers. One-minute EGF maps were recorded from standardized biatrial basket positions. Patients with source activity ≥26.5% were randomized 1:1 to PVI + EGF-guided ablation vs PVI only; patients without sources ≥26.5% threshold were not randomized. Follow-up and electrocardiographic monitoring occurred at 3, 6, and 12 months.
Results: We enrolled 85 patients (age 65.6 ± 9.3 years, 37% female, 24% long-standing PerAF). Thirty-four (40%) patients had no sources greater than threshold; at least 1 source greater than threshold was present in 46 (60%) (EGF-guided ablation, n = 22; control group, n = 26). Patients with sources were older (68.2 vs 62.6 years; P = 0.005) with higher CHADS-VASc scores (2.8 vs 1.9; P = 0.001). The freedom from safety events was 97.2%, and 95% of EGF-identified sources were successfully ablated. In randomized patients, AF-free survival at 12 months was 68% for EGF-guided ablation vs 17% for the control group (P = 0.042); freedom from AF/atrial tachycardia/atrial flutter at 12 months was 51% vs 14% (P = 0.103), respectively.
Conclusions: In nonparoxysmal AF patients undergoing redo ablation, EGF mapping identified AF sources in 60% of patients, and could be successfully ablated in 95%. Compared with PVI alone, PVI + source ablation improved AF-free survival by 51% on an absolute basis. (FLOW-AF: A Study to Evaluate the Ablacon Electrographic FLOW EGF Technology [A Randomized Controlled Study to Evaluate the Reliability of the Ablacon Electrographic FLOW (EGF) Algorithm Technology (Ablamap Software) to Identify AF Sources and Guide Ablation Therapy in Patients With Persistent Atrial Fibrillation]; NCT04473963).
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jacep.2024.03.040 | DOI Listing |
J Interv Card Electrophysiol
December 2024
Section of Electrophysiology, Division of Cardiology, Department of Medicine, Rush University Medical Center, 1717 W Congress Pkwy/Suite 317 Kellogg, Chicago, IL, 60612, USA.
Background: Catheter ablation is a key treatment for atrial fibrillation (AF), with high-power, very high-power short-duration and pulsed field ablation (PFA) being efficient options. However, direct comparisons between these techniques are lacking.
Objective: We performed a systematic review and meta-analysis, which included predominantly observational studies (four retrospective and one prospective study), to compare PFA and High-power short-duration (HPSD) and very high-power short-duration (vHPSD) radiofrequency (RF) ablation in patients with AF.
Eur Heart J Open
January 2025
Institute of Health Informatics Research, University College London, 222 Euston Road, London NW1 2DA, UK.
Aims: Cavotricuspid isthmus (CTI) ablation is the current ablation treatment for typical atrial flutter (AFL). However, post-ablation atrial tachyarrhythmias, mostly in the form of atrial fibrillation (AF), are frequently observed after CTI ablation. We aimed to evaluate the effectiveness and safety of concomitant or isolated pulmonary vein isolation (PVI) in patients with typical AFL scheduled for ablation.
View Article and Find Full Text PDFJ Cardiovasc Electrophysiol
December 2024
Department of Cardiovascular Medicine, Institute of Science Tokyo, Tokyo, Japan.
Background: Cryoballoon ablation is less operator-dependent than radiofrequency ablation. Recently, size-adjustable cryoballoons (SA-CBs) have become available. We sought to analyze the individual baseline operator experience's impact on procedural results.
View Article and Find Full Text PDFHeart Rhythm
December 2024
Heart Rhythm Management Centre, Postgraduate Program in Cardiac Electrophysiology and Pacing, Universitair Ziekenhuis Brussel-Vrije Universiteit Brussel, Brussels, Belgium.
Background: A novel focal lattice-tip catheter allowing the delivery of either pulsed field (PF) or radiofrequency (RF) energy has recently received regulatory approval. The technology features a proprietary 3-dimensional electroanatomic mapping system.
Objective: We describe the first real-world and multicenter experience.
Introduction: Pulsed field ablation (PFA) is a form of nonthermal energy that has been recently introduced for pulmonary vein isolation (PVI). A multi-electrode pentaspline catheter for delivery of PFA guided by fluoroscopy has become widely available for clinical use.
Methods And Results: In this study, we aimed to assess whether the addition of electroanatomical mapping (EAM) for confirmation of PVI in the acute phase can increase the efficacy of the procedure in terms of arrhythmia recurrences.