This study analyzed the stress distributions on zygomatic and dental implants placed in the zygomatic bone, supporting bones, and superstructures under occlusal loads after maxillary reconstruction with obturator prostheses. A total of 12 scenarios of 3-dimensional finite element models were constructed based on computerized tomography scans of a hemimaxillectomy patient. Two obturator prostheses were analyzed for each model. A total force of 600 N was applied from the palatal to buccal bones at an angle of 45°. The maximum and minimum principal stress values for bone and von Mises stress values for dental implants and prostheses were calculated. When zygomatic implants were applied to the defect area, the maximum principal stresses were similar in intensity to the other models; however, the minimum principal stress values were higher than in scenarios without zygomatic implants. In models that used zygomatic implants in the defect area, von Mises stress levels were significantly higher in zygomatic implants than in dental implants. In scenarios where the prosthesis was supported by tissue in the nondefect area, the maximum and minimum principal stress values on cortical bone were higher than in scenarios where implants were applied to defect and nondefect areas. In patients who lack an alveolar crest after maxillectomy, a custom bar-retained prosthesis placed on the dental implant should reduce stress on the zygomatic bone. The stress was higher on zygomatic implants without alveolar crest support than on dental implants.
Download full-text PDF |