Performance of a Wearable Ring in Controlled Hypoxia: A Prospective Observational Study.

JMIR Form Res

Movano Inc dba Movano Health, Pleasanton, CA, United States.

Published: June 2024

Background: Over recent years, technological advances in wearables have allowed for continuous home monitoring of heart rate and oxygen saturation. These devices have primarily been used for sports and general wellness and may not be suitable for medical decision-making, especially in saturations below 90% and in patients with dark skin color. Wearable clinical-grade saturation of peripheral oxygen (SpO) monitoring can be of great value to patients with chronic diseases, enabling them and their clinicians to better manage their condition with reliable real-time and trend data.

Objective: This study aimed to determine the SpO accuracy of a wearable ring pulse oximeter compared with arterial oxygen saturation (SaO) in a controlled hypoxia study based on the International Organization for Standardization (ISO) 80601-2-61:2019 standard over the range of 70%-100% SaO in volunteers with a broad range of skin color (Fitzpatrick I to VI) during nonmotion conditions. In parallel, accuracy was compared with a calibrated clinical-grade reference pulse oximeter (Masimo Radical-7). Acceptable medical device accuracy was defined as a maximum of 4% root mean square error (RMSE) per the ISO 80601-2-61 standard and a maximum of 3.5% RMSE per the US Food and Drug Administration guidance.

Methods: We performed a single-center, blinded hypoxia study of the test device in 11 healthy volunteers at the Hypoxia Research Laboratory, University of California at San Francisco, under the direction of Philip Bickler, MD, PhD, and John Feiner, MD. Each volunteer was connected to a breathing apparatus for the administration of a hypoxic gas mixture. To facilitate frequent blood gas sampling, a radial arterial cannula was placed on either wrist of each participant. One test device was placed on the index finger and another test device was placed on the fingertip. SaO analysis was performed using an ABL-90 multi-wavelength oximeter.

Results: For the 11 participants included in the analysis, there were 236, 258, and 313 SaO-SpO data pairs for the test device placed on the finger, the test device placed on the fingertip, and the reference device, respectively. The RMSE of the test device for all participants was 2.1% for either finger or fingertip placement, while the Masimo Radical-7 reference pulse oximeter RMSE was 2.8%, exceeding the standard (4% or less) and the Food and Drug Administration guidance (3.5% or less). Accuracy of SaO-SpO paired data from the 4 participants with dark skin in the study was separately analyzed for both test device placements and the reference device. The test and reference devices exceeded the minimum accuracy requirements for a medical device with RMSE at 1.8% (finger) and 1.6% (fingertip) and for the reference device at 2.9%.

Conclusions: The wearable ring meets an acceptable standard of accuracy for clinical-grade SpO under nonmotion conditions without regard to skin color.

Trial Registration: ClinicalTrials.gov NCT05920278; https://clinicaltrials.gov/study/NCT05920278.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11187508PMC
http://dx.doi.org/10.2196/54256DOI Listing

Publication Analysis

Top Keywords

test device
28
wearable ring
12
pulse oximeter
12
device
12
reference device
12
controlled hypoxia
8
oxygen saturation
8
dark skin
8
skin color
8
hypoxia study
8

Similar Publications

HIV prevention among adolescents travestis and transgender women in three Brazilian capitals, 2019-2023.

Epidemiol Serv Saude

January 2025

Universidade Federal da Bahia, Instituto de Saúde Coletiva, Salvador, BA, Brazil.

Objective: To describe HIV prevention strategies and gender-based discrimination among adolescent travestis and transgender women.

Methods: This was a cross-sectional study involving 148 adolescent travestis and transgender women aged 15 to 19 years in Salvador, Bahia state, São Paulo, São Paulo state, and Belo Horizonte, Minas Gerais state, conducted between February 2019 and March 2023. Fisher's exact test was performed to assess differences between prevention strategies and gender-based discrimination within healthcare services.

View Article and Find Full Text PDF

Dimensional engineering of interlayer for efficient large-area perovskite solar cells with high stability under ISOS-L-3 aging test.

Sci Adv

January 2025

Fujian Key Laboratory of Semiconductor Materials and Applications, CI Center for OSED, Department of Physics, Xiamen University, Xiamen 361005, P. R. China.

The utilization of low-dimensional perovskites (LDPs) as interlayers on three-dimensional (3D) perovskites has been regarded as an efficient strategy to enhance the performance of perovskite solar cells. Yet, the formation mechanism of LDPs and their impacts on the device performance remain elusive. Herein, we use dimensional engineering to facilitate the controllable growth of 1D and 2D structures on 3D perovskites.

View Article and Find Full Text PDF

Using time as an additional design parameter in electromagnetism, photonics, and wave physics is attracting considerable research interest, motivated by the possibility to explore physical phenomena and engineering opportunities beyond the limits of time-invariant systems. Here, we report the experimental demonstration of enhanced broadband absorption of electromagnetic waves in a continuously modulated time-varying system, exceeding one of the key theoretical limits of linear time-invariant absorbers. This is achieved by harnessing the frequency-wave vector transitions and enhanced interference effects enabled by breaking both continuous space- and time-translation symmetries in a periodically time-modulated absorbing structure operating at radio frequencies.

View Article and Find Full Text PDF

While piezoelectric sensing and energy-harvesting devices still largely rely on inorganic components, biocompatible and biodegradable piezoelectric materials, such as cellulose nanocrystals, might constitute optimal and sustainable building blocks for a variety of applications in electronics and transient implants. To this aim, however, effective methods are needed to position cellulose nanocrystals in large and high-performance architectures. Here, we report on scalable assemblies of cellulose nanocrystals in multilayered piezoelectric systems with exceptional response, for various application scopes.

View Article and Find Full Text PDF

Despite the advances in bionic reconstruction of missing limbs, the control of robotic limbs is still limited and, in most cases, not felt to be as natural by users. In this study, we introduce a control approach that combines robotic design based on postural synergies and neural decoding of synergistic behavior of spinal motoneurons. We developed a soft prosthetic hand with two degrees of actuation that realizes postures in a two-dimensional linear manifold generated by two postural synergies.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!