Background: Dried blood spot (DBS) sampling on cellulose cards suffers from varying blood haematocrit levels and from chromatographic effects, which have a direct impact on quantitative DBS analyses. Commercial volumetric microsampling devices were, therefore, introduced to mitigate these effects, however, these devices are not compatible with automated DBS processing systems and must be processed manually.
Results: Capillary electrophoresis (CE) instruments use fused-silica (FS) capillaries for precise and accurate liquid handling as well as for injection, separation, and quantitative analyses of liquid samples. These inherent features of an Agilent 7100 CE instrument were employed for the automated processing (elution and homogenization) of DBSs collected by hemaPEN® volumetric devices (2.74 μL of capillary blood per spot). The hemaPEN® samples were processed directly in CE vials by consecutive transfers of 56 μL of methanol and 14 μL of deionized water through the FS capillary in a sequence of 39 DBSs with repeatability of the liquid transfers better than 1.4 %. The resulting DBS eluates were homogenized by a quick air flush through the capillary and analyzed by the same capillary and CE instrument. Creatinine was selected as a clinically relevant model analyte and its endogenous concentrations in DBSs were determined by CE with capacitively coupled contactless conductivity detection (CE-CD) in a background electrolyte solution consisting of 50 mM acetic acid and 0.1 % (v/v) Tween 20 (pH 3.0). The overall repeatability of the automated DBS processing and CE-CD analyses of 39 DBSs was ≤7.1 % (peak areas) and ≤0.6 % (migration times), the calibration curve was linear in the 25-500 μM range (R = 0.9993) and covered all endogenous blood creatinine levels, the limit of detection was 5.0 μM, and sample throughput was >12 DBSs per hour. DBS ageing for 60 days and varying blood haematocrit levels (20-70 %) did not affect creatinine quantitative results (≤6.9 % for peak areas). Inter-capillary and inter-instrument repeatability was ≤7.7 % (peak areas) and ≤3.4 % (migration times) and demonstrated an excellent transferability of the proposed analytical concept among laboratories.
Significance And Novelty: This contribution is the first-ever report on the use of a single off-the-shelf analytical instrument for fully automated analyses of DBSs collected by commercial volumetric microsampling devices and holds great promise for future unmanned quantitative DBS analyses.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.aca.2024.342718 | DOI Listing |
J Pharm Biomed Anal
December 2024
Department of Clinical Pharmacy and Toxicology, Leiden University Medical Center, Leiden, the Netherlands; Leiden Network for Personalized Medicine, Leiden, the Netherlands. Electronic address:
Voclosporin is a potent immunosuppressive agent currently approved for treating active lupus nephritis. Based on its potential antiviral activity, it has also been investigated as immunosuppressive agent in an investigator-initiated study in SARS-CoV2 positive kidney transplant recipients. As with many immunosuppressive agents, optimizing dosing regimens to achieve therapeutic efficacy while minimizing toxicity remains a critical challenge in clinical practice.
View Article and Find Full Text PDFBMC Infect Dis
December 2024
Bill & Melinda Gates Medical Research Institute, Cambridge, MA, USA.
Background: Virus neutralising antibodies in serum are considered key correlates of protection for vaccines and monoclonal antibodies against respiratory syncytial virus (RSV). RSM01 is a novel, highly-potent, half-life-extended and fully-human monoclonal antibody candidate targeting the RSV prefusion F protein. Currently in Phase 1 development, RSM01 is primarily being developed to potentially provide an effective and affordable RSV prevention strategy in low- and middle-income countries.
View Article and Find Full Text PDFJ Chromatogr B Analyt Technol Biomed Life Sci
December 2024
Department of Pharmaceutical and Medical Chemistry, Clinical Pharmacy, University of Muenster, Muenster, Germany. Electronic address:
The number of prescriptions for new direct oral anticoagulants (DOACs) apixaban, edoxaban, rivaroxaban and dabigatran has increased exponentially in recent years, increasingly replacing the old gold standard, vitamin-K-antagonists. Due to their wide therapeutic range, therapeutic drug monitoring (TDM) is not required, although it has been proven that this could significantly reduce side effects. In order to develop a cost-efficient and simple method for the simultaneous detection of the DOACs and phenprocoumon, a new technology for sample preparation from capillary blood in the ambulant sector named VAMS® was integrated and an LC-MS detector with on-line solid phase extraction (SPE) applying a Turboflow HTLC Cyclone 1.
View Article and Find Full Text PDFAnal Chim Acta
January 2025
Department of Pharmaceutical Chemistry and Pharmaceutical Analysis, Faculty of Pharmacy in Hradec Králové, Charles University, Akademika Heyrovského 1203, 500 03, Hradec Králové, Czech Republic. Electronic address:
Background: Microsampling of biological fluids followed by innovative sample pre-treatment reflects trends in bioanalytical chemistry. Volumetric absorptive microsampling (VAMS) enables exact whole blood volume collection and reduces the impact of hematocrit on the assay. In animal studies, it complies with the 3R principles (refine, reduce, replace).
View Article and Find Full Text PDFAnal Bioanal Chem
January 2025
Laboratoire de Toxicologie, Fédération de Toxicologie, Hôpital Lariboisière AP-HP, 2 Rue Ambroise Paré, 75010, Paris, France.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!