AI Article Synopsis

  • Dried blood spot (DBS) sampling faces challenges like varying haematocrit levels and chromatographic issues, prompting the use of commercial volumetric microsampling devices, which unfortunately require manual processing and aren't automated.
  • The study utilized an Agilent 7100 capillary electrophoresis (CE) instrument for automated processing of DBS samples collected via the hemaPEN® device, achieving precise liquid handling and repeatable results with blood sample analyses.
  • The analysis focused on creatinine levels, showing an excellent calibration curve and high throughput of over 12 DBSs per hour, while demonstrating that factors like sample aging and blood haematocrit levels did not significantly impact the results.

Article Abstract

Background: Dried blood spot (DBS) sampling on cellulose cards suffers from varying blood haematocrit levels and from chromatographic effects, which have a direct impact on quantitative DBS analyses. Commercial volumetric microsampling devices were, therefore, introduced to mitigate these effects, however, these devices are not compatible with automated DBS processing systems and must be processed manually.

Results: Capillary electrophoresis (CE) instruments use fused-silica (FS) capillaries for precise and accurate liquid handling as well as for injection, separation, and quantitative analyses of liquid samples. These inherent features of an Agilent 7100 CE instrument were employed for the automated processing (elution and homogenization) of DBSs collected by hemaPEN® volumetric devices (2.74 μL of capillary blood per spot). The hemaPEN® samples were processed directly in CE vials by consecutive transfers of 56 μL of methanol and 14 μL of deionized water through the FS capillary in a sequence of 39 DBSs with repeatability of the liquid transfers better than 1.4 %. The resulting DBS eluates were homogenized by a quick air flush through the capillary and analyzed by the same capillary and CE instrument. Creatinine was selected as a clinically relevant model analyte and its endogenous concentrations in DBSs were determined by CE with capacitively coupled contactless conductivity detection (CE-CD) in a background electrolyte solution consisting of 50 mM acetic acid and 0.1 % (v/v) Tween 20 (pH 3.0). The overall repeatability of the automated DBS processing and CE-CD analyses of 39 DBSs was ≤7.1 % (peak areas) and ≤0.6 % (migration times), the calibration curve was linear in the 25-500 μM range (R = 0.9993) and covered all endogenous blood creatinine levels, the limit of detection was 5.0 μM, and sample throughput was >12 DBSs per hour. DBS ageing for 60 days and varying blood haematocrit levels (20-70 %) did not affect creatinine quantitative results (≤6.9 % for peak areas). Inter-capillary and inter-instrument repeatability was ≤7.7 % (peak areas) and ≤3.4 % (migration times) and demonstrated an excellent transferability of the proposed analytical concept among laboratories.

Significance And Novelty: This contribution is the first-ever report on the use of a single off-the-shelf analytical instrument for fully automated analyses of DBSs collected by commercial volumetric microsampling devices and holds great promise for future unmanned quantitative DBS analyses.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.aca.2024.342718DOI Listing

Publication Analysis

Top Keywords

volumetric microsampling
12
microsampling devices
12
peak areas
12
automated analyses
8
dried blood
8
blood spot
8
varying blood
8
blood haematocrit
8
haematocrit levels
8
quantitative dbs
8

Similar Publications

Dried blood spot LC-MS/MS quantification of voclosporin in renal transplant recipients using volumetric dried blood spot sampling.

J Pharm Biomed Anal

December 2024

Department of Clinical Pharmacy and Toxicology, Leiden University Medical Center, Leiden, the Netherlands; Leiden Network for Personalized Medicine, Leiden, the Netherlands. Electronic address:

Voclosporin is a potent immunosuppressive agent currently approved for treating active lupus nephritis. Based on its potential antiviral activity, it has also been investigated as immunosuppressive agent in an investigator-initiated study in SARS-CoV2 positive kidney transplant recipients. As with many immunosuppressive agents, optimizing dosing regimens to achieve therapeutic efficacy while minimizing toxicity remains a critical challenge in clinical practice.

View Article and Find Full Text PDF

Background: Virus neutralising antibodies in serum are considered key correlates of protection for vaccines and monoclonal antibodies against respiratory syncytial virus (RSV). RSM01 is a novel, highly-potent, half-life-extended and fully-human monoclonal antibody candidate targeting the RSV prefusion F protein. Currently in Phase 1 development, RSM01 is primarily being developed to potentially provide an effective and affordable RSV prevention strategy in low- and middle-income countries.

View Article and Find Full Text PDF

Development and validation of a quantification method for direct oral anticoagulants from capillary blood using volumetric absorptive microsampling and online SPE-LC-MS.

J Chromatogr B Analyt Technol Biomed Life Sci

December 2024

Department of Pharmaceutical and Medical Chemistry, Clinical Pharmacy, University of Muenster, Muenster, Germany. Electronic address:

The number of prescriptions for new direct oral anticoagulants (DOACs) apixaban, edoxaban, rivaroxaban and dabigatran has increased exponentially in recent years, increasingly replacing the old gold standard, vitamin-K-antagonists. Due to their wide therapeutic range, therapeutic drug monitoring (TDM) is not required, although it has been proven that this could significantly reduce side effects. In order to develop a cost-efficient and simple method for the simultaneous detection of the DOACs and phenprocoumon, a new technology for sample preparation from capillary blood in the ambulant sector named VAMS® was integrated and an LC-MS detector with on-line solid phase extraction (SPE) applying a Turboflow HTLC Cyclone 1.

View Article and Find Full Text PDF

Volumetric absorptive microsampling meets electromembrane extraction for the first time: Case example of doxorubicin and its metabolite in whole blood samples.

Anal Chim Acta

January 2025

Department of Pharmaceutical Chemistry and Pharmaceutical Analysis, Faculty of Pharmacy in Hradec Králové, Charles University, Akademika Heyrovského 1203, 500 03, Hradec Králové, Czech Republic. Electronic address:

Background: Microsampling of biological fluids followed by innovative sample pre-treatment reflects trends in bioanalytical chemistry. Volumetric absorptive microsampling (VAMS) enables exact whole blood volume collection and reduces the impact of hematocrit on the assay. In animal studies, it complies with the 3R principles (refine, reduce, replace).

View Article and Find Full Text PDF

Quantitative analysis of cannabinoids and metabolites in oral fluid by volumetric absorptive microsampling combined with UHPLC-HRMS.

Anal Bioanal Chem

January 2025

Laboratoire de Toxicologie, Fédération de Toxicologie, Hôpital Lariboisière AP-HP, 2 Rue Ambroise Paré, 75010, Paris, France.

Article Synopsis
  • The increasing legalization of cannabis worldwide has led to a growing need for methods to accurately identify and quantify cannabis consumption, especially through non-invasive means.
  • Oral fluid, as an alternative to traditional blood and urine tests, offers advantages such as easier self-sampling and reduced risk of tampering, although it typically requires larger volumes to test effectively.
  • The study presents a new method that allows for the quantification of seven cannabinoids from just 20 µL of oral fluid using a specialized sampling device and advanced chemical techniques, achieving precise detection limits suitable for legal and clinical contexts.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!