Leadless pacemaker implementation at the right atrial appendage apex: An initial preclinical assessment.

Pacing Clin Electrophysiol

Division of Cardiology, Department of Internal Medicine, Chang Gung Memorial Hospital, Linkou, Taoyuan, Taiwan; College of Medicine, Chang Gung University, Taoyuan, Taiwan., Chang Gung Memorial Hospital, Taoyuan, Taiwan.

Published: August 2024

Objective: This study evaluates the feasibility and efficacy of implanting a leadless pacemaker at the right atrial appendage (RAA) in a preclinical minipig model, aiming to address the limitations of atrial pacing with current leadless devices like the Medtronic Micra, which is typically used for right ventricular implantation.

Methods: Four minipigs, each with a median body weight of 45.8 ± 10.0 kg, underwent placement of the Micra transcatheter pacing system (TPS) via the right femoral vein into the RAA apex. The pacing performance was assessed over 1-week (short-term) and 3-month (long-term) periods.

Outcomes: The initial findings indicated successful implantation, with satisfactory intrinsic R-wave amplitudes and pacing threshold. In the following period, the sensitivity, threshold, and impedance were stable with time. Notably, upon explanation at 3 months, a deep myocardial penetration by the device was observed, necessitating a redesign for safe long-term use in a growing subject's heart.

Conclusion: While initial results suggest that RAA apex placement of the Micra TPS is promising for potential inclusion in a dual-chamber pacing system, the issue of myocardial penetration highlights the need for device redesign to ensure safety and effectiveness in long-term applications.

Download full-text PDF

Source
http://dx.doi.org/10.1111/pace.15003DOI Listing

Publication Analysis

Top Keywords

leadless pacemaker
8
atrial appendage
8
placement micra
8
pacing system
8
raa apex
8
myocardial penetration
8
pacing
5
pacemaker implementation
4
implementation atrial
4
appendage apex
4

Similar Publications

Minimizing for Maximum Benefit: An Illustrative Case-Series of Atrial Only Leadless Pacing.

Pacing Clin Electrophysiol

December 2024

Division of Cardiology, Department of Medicine, Weill Cornell Medicine, New York Presbyterian Hospital, New York, USA.

Leadless pacing technology now includesdedicated atrial helix-fixation leadless pacemakers (LPs), expanding theapplication of leadless devices for patients with sinus node dysfunction andatrioventricular block during sinus rhythm. This first reportedcase-series of atrial LPs describes and discusses the potential use-casescenarios of recently approved helix-fixation atrial LPs. The article highlights important concepts regarding their use, including implantationtechniques, programming, battery conservation, and the low rate of progressionof AV block in patients implanted with AAI(R) pacemakers.

View Article and Find Full Text PDF

Reel's syndrome (RS) is an unusual cause of pacemaker lead dislodgement. We present the case of a 59-year-old female patient with Down syndrome (DS) implanted with a dual-chamber endovascular pacemaker due to symptomatic sinus node disfunction, reporting several syncopal episodes in last days and showing abnormal electrical parameters at the 2-months follow-up due to RS. The malfunctioning device was removed and an endocardial leadless pacing system was implanted.

View Article and Find Full Text PDF

Background: Delayed lead perforation is a rare complication of cardiac implantable electronic device (CIED). Clinical presentations range from completely asymptomatic to pericardial tamponade. Surgical lead extraction is recommended and transvenous lead extraction (TLE) with surgical backup is an alternative method.

View Article and Find Full Text PDF

Cardiac Implantable Electronic Devices in Cardiac Transplant Patients: A Comprehensive Review.

Cardiol Rev

November 2024

Department of Cardiovascular Medicine, Cardiovascular Analytics Group, Islamabad, Pakistan.

A fraction of patients (approximately 10%) undergoing heart transplantation require permanent pacemaker (PPM) implantation due to sinus node dysfunction or atrioventricular block, occurring either shortly after surgery or later. The incidence of PPM implantation has declined to less than 5% with the introduction of bicaval anastomosis transplantation surgery. Pacing dependency during follow-up varies among recipients.

View Article and Find Full Text PDF

Unlabelled: New cardiac implantable electronic devices (CIEDs), such as leadless pacemakers and subcutaneous implantable cardioverter defibrillators (S-ICDs), are being used in patients with adult congenital heart disease. The selection of CIEDs often requires careful consideration due to technical challenges posed by a unique heart structure. A 27-year-old man following a surgical tetralogy of Fallot (TOF) repair developed non-sustained ventricular tachycardia, sick sinus syndrome, and complete atrioventricular block.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!