Introduction: Information on tissue perfusion in the foot is important when treating patients with chronic limb-threatening ischemia. This study aims to test the reliability of different magnetic resonance sequences when measuring perfusion in the foot.
Methods: Sixteen healthy volunteers had their right foot scanned in a test/retest study with six different magnetic resonance sequences (BOLD, multi-echo gradient echo (mGRE), 2D and 3D pCASL, PASL FAIR, and DWI with intravoxel incoherent motion (IVIM) with quantitative measurements of perfusion. For five sequences, cuff-induced ischemia followed by a hyperactive response was measured. Images of the feet were segmented into angiosomes and perfusion data were extracted from the five angiosomes.
Results: BOLD, PASL FAIR, mGRE, and DWI with IVIM had low mean differences between the first and second scans, while the results of 2D and 3D pCASL had the highest differences. Based on a paired t-test, BOLD, and FAIR were able to distinguish between perfusion and no perfusion in all angiosomes with p-values below 0.01. This was not the case with 2D and 3D pCASL with p-values above 0.05 in all angiosomes. The mGRE could not distinguish between perfusion and no perfusion in the lateral side of the foot.
Conclusion: BOLD, mGRE, pASL FAIR, and DWI with IVIM seem to give more robust results compared to 2D and 3D pCASL. Further studies on patients with peripheral artery disease should explore if the sequences can have clinical relevance when assessing tissue ischemia and results of revascularization.
Implications For Practice: This study provides knowledge that could be used to improve the diagnosis of patient with chronic limb-threatening ischemia to explore tissue perfusion.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.radi.2024.05.007 | DOI Listing |
Radiography (Lond)
July 2024
Department of Radiology, Kolding Hospital, Kolding, Denmark; Department of Regional Health Research, University of Southern Denmark, Odense, Denmark; Health Sciences Research Centre, UCL University College, Odense M, Denmark. Electronic address:
Introduction: Information on tissue perfusion in the foot is important when treating patients with chronic limb-threatening ischemia. This study aims to test the reliability of different magnetic resonance sequences when measuring perfusion in the foot.
Methods: Sixteen healthy volunteers had their right foot scanned in a test/retest study with six different magnetic resonance sequences (BOLD, multi-echo gradient echo (mGRE), 2D and 3D pCASL, PASL FAIR, and DWI with intravoxel incoherent motion (IVIM) with quantitative measurements of perfusion.
Front Physiol
January 2024
Krembil Brain Institute, University Health Network, Toronto, ON, Canada.
Cerebral blood flow (CBF) is a critical physiological parameter of brain health, and it can be non-invasively measured with arterial spin labeling (ASL) MRI. In this study, we evaluated and optimized whole-brain, high-resolution ASL as an alternative to the low-resolution ASL employed in the routine assessment of CBF in both healthy participants and patients. Two high-resolution protocols (i.
View Article and Find Full Text PDFEur Radiol
January 2019
Medical Research Council (MRC), London Institute of Medical Sciences (LMS), Imperial College London, Hammersmith Hospital Campus, Du Cane Road, London, W12 0NN, UK.
Objectives: We investigated the feasibility and reproducibility of free-breathing motion-corrected multiple inversion time (multi-TI) pulsed renal arterial spin labelling (PASL), with general kinetic model parametric mapping, to simultaneously quantify renal perfusion (RBF), bolus arrival time (BAT) and tissue T.
Methods: In a study approved by the Health Research Authority, 12 healthy volunteers (mean age, 27.6 ± 18.
Diagnostics (Basel)
June 2018
Sir Peter Mansfield Imaging Centre, School of Physics and Astronomy, University of Nottingham, Nottingham NG7 2RD, UK.
A number of imaging readout schemes are proposed for renal arterial spin labeling (ASL) to quantify kidney cortex perfusion, including gradient echo-based methods of balanced fast field echo (bFFE) and gradient-echo echo-planar imaging (GE-EPI), or spin echo-based schemes of spin-echo echo-planar imaging (SE-EPI) and turbo spin-echo (TSE). Here, we compare these two-dimensional (2D) imaging schemes to evaluate the optimal imaging scheme for pulsed ASL (PASL) assessment of human kidney cortex perfusion at 3 T. Ten healthy volunteers with normal renal function were scanned using each 2D multi-slice imaging scheme, in combination with a respiratory triggered flow-sensitive alternating inversion recovery (FAIR) ASL scheme on a 3 T Philips Achieva scanner.
View Article and Find Full Text PDFMagn Reson Med
December 2013
Helen Wills Institute of Neuroscience, University of California, Berkeley, Advanced MRI Technologies, Sebastopol, California, USA.
Purpose: Simultaneous multi-slice (SMS) echo planar imaging (EPI) is incorporated into two-dimensional (2D) arterial spin labeling (ASL) imaging to produce more slices for measuring perfusion in a larger region of the brain than currently possible with multi-slice EPI.
Methods: Pulsed ASL (PASL) preparations using FAIR and QUIPSS II techniques were combined with SMS-EPI. Testing was performed in four subjects at 3 Tesla.