Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Continuous C8 stationary phase gradients are created on commercial Waters Symmetry Shield RP8 columns by strategically cleaving the C8 moieties in a time-dependent fashion. The method relies on the controlled infusion of a trifluoroacetic acid/water/acetonitrile solution through the column to cleave the organic functionality (e.g., C8) from the siloxane framework. The bond cleavage solution is reactive enough to cleave the functional groups, even with polar groups embedded within the stationary phase to protect the silica. Both the longitudinal and radial heterogeneity were evaluated by extruding the silica powder into polyethylene tubing and evaluating the percent carbon content in the different sections using thermogravimetric analysis (TGA). TGA analysis shows the presence of a stationary phase gradient in the longitudinal direction but not in the radial direction. Two different gradient profiles were formed with good reproducibility by modifying the infusion method: one exhibited an 'S'-shaped gradient while the other exhibited a steep exponential-like gradient. The gradients were characterized chromatographically using test mixtures, and the results showed varied retention characteristics and an enhanced ability to resolve nicotine analytes.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.chroma.2024.464974 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!