  • The study compares the peak insertion torques of three types of screws used for cervical spine fixation: lateral mass screws (LMS), pedicle screws, and the new Attallah screw.
  • Results indicate that while the Attallah screw has comparable torque to pedicle screws at certain cervical levels, it generally shows lower strength at C5 and C6.
  • Overall, the findings suggest that the Attallah screw is a viable option for posterior cervical fixation, but pedicle screws are still considered superior in strength, especially between C5 and C6.

Article Abstract

Posterior fixation of the subaxial cervical spine (SCS) commonly relies on the application of lateral mass screws (LMS), with pedicle screws being a less prevalent alternative. The present study provides another option: A recently introduced novel approach, the Attallah screw, intended to ensure a safety profile comparable to that of LMS, combined with a strength profile similar to that of pedicle screws. The focus of the present study is the comparative analysis of peak insertion torques for these three screw types. Employing standard surgical techniques and instruments, Attallah screws were scheduled for insertion on the right side of the SCS in 15 cadavers, pedicle screws on the left side in 8 cadavers, and LMS on the left side in the remaining 7 cadavers. The peak insertion torque was recorded using an electronic torque screwdriver. The results revealed that the peak insertion torques were similar in the pedicle and the Attallah screw at C3, C4 and C7, but differed at C5 (mean ± SD; pedicle, 79.5±19.6 cNm; Attallah, 56.7±18.5 cNm; P=0.029) and C6 (pedicle, 85.4±28.7 cNm; Attallah, 49.8±17.9 cNm; P=0.004) in favor of the superior pedicle screw measurements. The peak insertion torques of the pedicle screw were superior to the corresponding data from the LMS from C4 to C7. By contrast, the peak insertion torques of the Attallah screw were only superior to those of the LMS at C7 (Attallah, 69.5±24.5 cNm; lateral mass, 40.5±21.4 cNm; P=0.030), although similar trends were observed at the other cervical levels. On the whole, the findings presented herein indicate the level-dependent superior robustness of the Attallah screw as a posterior cervical fixation method compared to the LMS. However, from a biomechanical perspective, the pedicle screw remains the preeminent choice for fixation within the C5-C6 range.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11097134PMC
http://dx.doi.org/10.3892/mi.2024.159DOI Listing

Publication Analysis

Top Keywords

attallah screw
20
peak insertion
20
insertion torques
16
pedicle screws
12
pedicle screw
12
attallah
9
pedicle
9
subaxial cervical
8
lateral mass
8
screw
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!