In a series of 189 patients, 197 operations (251 stapled anastomoses) were performed. U.S. Surgical Corporation instruments (GIA n = 156, EEA n = 36, TA n = 59) were used in surgery of the stomach, pancreas, small and large bowel, and the rates of complications were investigated. The EEA gave 14% complications because of the frequent failure (33%) of the anastomosis after low anterior resection. The reasons for failure were incomplete rings, colonic wall damage and too short rectal stumps. The use of the EEA in esophago-jejunostomy and esophago-gastrostomy was safe and the postoperative course uneventful. The GIA was used in intestinal and colonic surgery and was pleasant and quick to use (right hemicolectomy: 1 hour 40 minutes; sigmoid resection: 1 hour 45 minutes). Complications occurred in the intestine (6%) and colon (7%). Only with GIA stapled anastomoses was there a risk of hemorrhage (2.5%). The incidence of anastomotic leakage (all staplers) was 3.6%; the stapler was responsible for a 2.0% mortality. Weighing up the indications and contraindications of stapling devices, this method presents an advantageous but not indispensable method in gastrointestinal surgery.
Download full-text PDF |
Source |
---|
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!