Measuring Visual Fields in Children With Glaucoma Using a Portable Tablet.

Transl Vis Sci Technol

Department of Optometry and Vision Sciences, University of Melbourne, Parkville, Victoria, Australia.

Published: May 2024

AI Article Synopsis

  • The study aimed to compare visual field testing results between an iPad app (Melbourne Rapid Fields) and the standard Humphrey Field Analyser in children with glaucoma.
  • Sixteen children aged 7 to 15 participated, and findings showed that the app's results closely matched the standard device, demonstrating high correlation and faster test times.
  • The children favored the iPad app over the standard test, suggesting it's an effective and preferable option for assessing visual fields in pediatric glaucoma cases.

Article Abstract

Purpose: To compare perimetric outcomes of an iPad perimetry app (Melbourne Rapid Fields [MRF]) with those of the Humphrey Field Analyser (HFA) testing children with glaucoma.

Methods: Sixteen children diagnosed and treated for glaucoma were recruited to evaluate their perimetric performance over two visits. At each visit, they undertook visual field assessment using the MRF application as well as the HFA. The HFA test was part of their usual clinical work up and a clinical assistant judged which test format (24-2 SITA standard or SITA fast) might be suited to the testing of that child. The primary outcome measure was the association and repeatability of mean deviation (MD) for the MRF and HFA tests, by way of regression, intraclass correlation coefficient and Bland-Altman analysis. Secondary measures were comparisons of pattern deviation indices, test times as well as an indication of participant test preference. Summary data show means ± standard deviation.

Results: The age for our cohort was 7 to 15 years of age (mean, 10.0 ± 2.4 years of age). The MRF MD was in close concordance to HFA MD with an intraclass correlation coefficient of 0.91 (95% confidence interval, 0.82-0.95). Bland-Altman analysis found little bias (-0.6 dB) and a 95% coefficient of repeatability of 2.1 dB in eyes having a normal HFA MD. In eyes with glaucomatous visual field defects the 95% coefficient of repeatability at retest was much larger for both the MRF (10.5 dB) as well as for the HFA (10.0 dB). Average MRF test times (5.6 ± 1.2 minutes) were similar to SITA Fast (5.4 ± 1.9 minutes) with both being significantly faster than SITA standard (8.6 ± 1.4 minutes; P < 0.001). All children chose testing with the MRF as their preference.

Conclusions: MRF correlated strongly with HFA and was preferred by the children over the HFA. MRF is suitable for perimetric evaluation of children with glaucoma.

Translational Relevance: This study finds that an iPad based visual field test can be used with children having glaucoma to yield outcomes similar to SITA-fast. Children indicate a preference for such testing.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11103736PMC
http://dx.doi.org/10.1167/tvst.13.5.10DOI Listing

Publication Analysis

Top Keywords

visual field
12
hfa
9
children
8
children glaucoma
8
mrf
8
well hfa
8
sita standard
8
sita fast
8
intraclass correlation
8
correlation coefficient
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!