Introduction: Intraoral scanners commonly used in orthodontic offices now offer near-infrared imaging (NIRI) technology, advertised as a screening tool to identify interproximal caries. This study aimed to evaluate the reliability and validity of NIRI detection of interproximal carious lesions in a common intraoral scanner (iTero Element 5D; Align Technology, San Jose, Calif) with and without bitewing radiograph complement, compared with a microcomputed tomography (micro-CT) reference standard.
Methods: Extracted human posterior teeth (premolars and molars) were selected for early (noncavitated) interproximal carious lesions (n = 39) and sound control surfaces (n = 47). The teeth were scanned via micro-CT for evaluation by 2 blinded evaluators using consensus scoring. The teeth were mounted to simulate anatomic interproximal contacts and underwent a NIRI scan using iTero Element 5D and bitewing radiographs. Two trained, calibrated examiners independently evaluated (1) near-infrared images alone with clinical photograph, (2) bitewing radiograph alone with clinical photograph, and (3) near-infrared images with bitewing radiograph and clinical photograph in combination, after at least a 10-day washout period between each evaluation.
Results: Interrater reliability was highest for NIRI alone (k = 0.533) compared with bitewing radiograph alone (k = 0.176) or in combination (k = 0.256). NIRI alone showed high specificity (0.83-0.96) and moderate sensitivity (0.42-0.63) compared with a micro-CT reference standard. Dentin lesions were significantly more reliably detected than enamel lesions.
Conclusions: After rigorous training and calibration, NIRI can be used with moderate reliability, high specificity, and moderate sensitivity to detect noncavitated interproximal carious lesions.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.ajodo.2024.03.013 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!