Repeatability and Agreement of 4 Biometers Measuring Corneal Astigmatism in Eyes With Irregular Corneal Astigmatism Component.

Am J Ophthalmol

From the National Clinical Research Center for Ocular Diseases, Eye Hospital (L.W., X.S., J.L.F., J.J.H., K.J.Z., R.L., A.Y.Y.), Wenzhou Medical University, Wenzhou, China; National Engineering Research Center of Ophthalmology and Optometry, Eye Hospital (A.Y.Y.), Wenzhou Medical University, Wenzhou, China. Electronic address:

Published: September 2024

Purpose: To investigate the repeatability and agreement of corneal astigmatism measurements in eyes with irregular corneal astigmatism component (ICAC) using four devices: IOLMaster 700 biometer, Lenstar 900 biometer, iTrace, and Pentacam.

Design: Prospective cross-sectional reliability analysis.

Methods: Sixty-four eyes (52 patients) with ICAC were examined three times using the four devices. The eye with ICAC in this study is defined as the cornea has a certain degree of irregular astigmatism (asymmetric and/or skewed bowtie pattern of corneal topography according to corneal topography classification), accompanied with total corneal higher-order aberrations in the 4 mm zone of 0.3 µm or greater. Corneal astigmatism was evaluated using three categories: anterior corneal astigmatism (ACA), posterior corneal astigmatism, and total corneal astigmatism (TCA). The repeatability was determined using the ∆Ast (arithmetic mean of vector differences among three repeated corneal astigmatism measurements). Bland-Altman plots and astigmatism vector analyses were employed to assess agreement.

Results: The IOLMaster 700 (∆Ast = 0.27 ± 0.20 D) showcased higher repeatability in ACA measurements compared to iTrace (∆Ast = 0.37 ± 0.38 D, P = .040) and Pentacam (∆Ast = 0.50 ± 0.22 D, P < .001), and paralleled the performance of Lenstar 900 (∆Ast = 0.31 ± 0.26 D, P = .338). The Pentacam (∆Ast = 0.09 ± 0.07 D, P < .001) demonstrated superior repeatability in posterior corneal astigmatism, whereas the IOLMaster 700 (∆Ast = 0.33 ± 0.23 D, P < .001) excelled in TCA. The IOLMaster 700 exhibited good agreement with either Lenstar 900 or iTrace, characterized by narrow 95% limits of agreement and clinically acceptable vector differences. Conversely, vector differences between Pentacam and the other three devices in ACA and TCA measurements were clinically significant, exceeding 0.50 D (all P < .05).

Conclusions: In terms of repeatability of corneal astigmatism measurements in eyes with ICAC, the IOLMaster 700 and Lenstar 900 outperformed iTrace and Pentacam. While the IOLMaster 700 can be used interchangeably with either Lenstar 900 or iTrace, the Pentacam is not interchangeable with the other three devices.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ajo.2024.05.002DOI Listing

Publication Analysis

Top Keywords

corneal astigmatism
44
iolmaster 700
24
lenstar 900
20
corneal
14
astigmatism
13
astigmatism measurements
12
vector differences
12
repeatability agreement
8
eyes irregular
8
irregular corneal
8

Similar Publications

Purpose: To evaluate the clinical outcomes of a new FDA approved toric aspheric hydrophobic acrylic intraocular lens.

Patients And Methods: This is a single surgeon, single-arm, on-label, non-randomized, prospective observational study. Thirty patients underwent bilateral cataract surgery (60 eyes) with placement of a Clareon™ Toric IOL in each eye.

View Article and Find Full Text PDF

Introduction: Virtual reality-based simulation is an educational tool that has been proven to increase participants' self-perceived, confidence, and skill. However, the use of VR is associated with virtual reality sickness (VRS). The purpose of this study is to determine related factors of VRS in an emergency setting simulation-based training, hence providing information and mitigation plan to enhance and optimize learning outcomes.

View Article and Find Full Text PDF

This prospective cohort study is aimed to investigate circadian variations in corneal parameters, focusing on sleep-deprived subjects. Sixty-four healthy individuals (age range: 21-76 years) actively participated in this study, undergoing examinations at least five times within a 24-hour timeframe. The analysis encompassed keratometric parameters of the cornea's front (F) and back (B) surfaces, refractive power in flattest and steepest axes (K1, K2), astigmatism (Astig) and its axis (Axis), aspheric coefficient (Asph), corneal pachymetry values of thinnest corneal thickness (Pachy Min) and corneal thickness in the center of the pupil (Pachy Pupil), volume relative to the 3 and 10 mm corneal diagonal (Vol D3, Vol D10) and surface variance index (ISV).

View Article and Find Full Text PDF

Corneal astigmatism after temporary discontinuation from long-term spherical and toric orthokeratology.

Cont Lens Anterior Eye

January 2025

Laboratory of Optometry and Vision Sciences, West China School of Medicine, Sichuan University, Chengdu, China; Department of Optometry and Vision Sciences, West China School of Medicine, Sichuan University, Chengdu, China.

Objective: To investigate changes in corneal astigmatism (curvatures and corneal sagittal height difference (C-sagDiff) at 8-mm chord) after temporary discontinuation from long-term spherical and toric orthokeratology (ortho-k) lens wear in a Chinese population.

Methods: Relevant data were retrieved from 110 patients (55 wearing spherical ortho-k lenses, 55 wearing toric ortho-k lenses) who have been undergoing ortho-k treatment for at least one year and stopped lens wear temporarily for scheduled lens replacement. The topographic and refraction data at baseline and post-discontinuation was collected for all the patients.

View Article and Find Full Text PDF

Objective: This study investigates the refractive accuracy of eight intraocular lens (IOL) power calculation formulas in patients with postoperative refractive surprise after phacoemulsification. It aims to determine if a different formula could result in better refractive outcomes in these eyes.

Methods And Analysis: We retrospectively reviewed consecutive patients undergoing uncomplicated phacoemulsification as a sole procedure between March 2007 and September 2020 at the University of Washington by glaucoma subspecialists as part of a study investigating cataract surgery in normal eyes.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!