Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Purpose: The purpose of this study was to compare micromotion of two new cementless tibial baseplates to a cementless design with well-published clinical success.
Methods: Three cementless tibial baseplate designs (fixed-bearing [FB] with keel and cruciform pegs, rotating-platform with porous central cone and pegs, FB with cruciform keel and scalloped pegs) were evaluated on sawbone models. Loading was applied to the baseplate at a rate of 1 Hz for 10,000 cycles, which represents 6-8 weeks of stair descent. This time frame also represents the approximate time length for the induction of biologic fixation of cementless implants. Compressive and shear micromotion at the sawbone-implant interface were measured.
Results: At the end of the loading protocol, the central cone rotating-platform design exhibited greater micromotion at the anterior (p < 0.001), posterior (p < 0.001) and medial locations (p = 0.049) compared to the other two implants. The central cone design also exhibited greater translational micromotion in the sagittal plane at the medial (p = 0.001) and lateral locations (p = 0.034) and in the coronal plane anteriorly (p = 0.007).
Conclusion: The cementless central cone rotating-platform baseplate demonstrated greater vertical and translational micromotion compared to the two FB baseplates with a keel underloading. This may indicate lower initial mechanical stability in implants without a keel, which possibly affects osseointegration. The implication of this is yet unknown and requires further long-term clinical follow-up to correlate these laboratory findings.
Level Of Evidence: V (biomechanical study).