Background: Preclinical low-count positron emission tomography (LC-PET) imaging offers numerous advantages such as facilitating imaging logistics, enabling longitudinal studies of long- and short-lived isotopes as well as increasing scanner throughput. However, LC-PET is characterized by reduced photon-count levels resulting in low signal-to-noise ratio (SNR), segmentation difficulties, and quantification uncertainties.

Purpose: We developed and evaluated a novel deep-learning (DL) architecture-Attention based Residual-Dilated Net (ARD-Net)-to generate standard-count PET (SC-PET) images from LC-PET images. The performance of the ARD-Net framework was evaluated for numerous low count realizations using fidelity-based qualitative metrics, task-based segmentation, and quantitative metrics.

Method: Patient Derived tumor Xenograft (PDX) with tumors implanted in the mammary fat-pad were subjected to preclinical [F]-Fluorodeoxyglucose (FDG)-PET/CT imaging. SC-PET images were derived from a 10 min static FDG-PET acquisition, 50 min post administration of FDG, and were resampled to generate four distinct LC-PET realizations corresponding to 10%, 5%, 1.6%, and 0.8% of SC-PET count-level. ARD-Net was trained and optimized using 48 preclinical FDG-PET datasets, while 16 datasets were utilized to assess performance. Further, the performance of ARD-Net was benchmarked against two leading DL-based methods (Residual UNet, RU-Net; and Dilated Network, D-Net) and non-DL methods (Non-Local Means, NLM; and Block Matching 3D Filtering, BM3D). The performance of the framework was evaluated using traditional fidelity-based image quality metrics such as Structural Similarity Index Metric (SSIM) and Normalized Root Mean Square Error (NRMSE), as well as human observer-based tumor segmentation performance (Dice Score and volume bias) and quantitative analysis of Standardized Uptake Value (SUV) measurements. Additionally, radiomics-derived features were utilized as a measure of quality assurance (QA) in comparison to true SC-PET. Finally, a performance ensemble score (EPS) was developed by integrating fidelity-based and task-based metrics. Concordance Correlation Coefficient (CCC) was utilized to determine concordance between measures. The non-parametric Friedman Test with Bonferroni correction was used to compare the performance of ARD-Net against benchmarked methods with significance at adjusted p-value ≤0.01.

Results: ARD-Net-generated SC-PET images exhibited significantly better (p ≤ 0.01 post Bonferroni correction) overall image fidelity scores in terms of SSIM and NRMSE at majority of photon-count levels compared to benchmarked DL and non-DL methods. In terms of task-based quantitative accuracy evaluated by SUV and SUV ARD-Net exhibited less than 5% median absolute bias for SUV compared to true SC-PET and lower degree of variability compared to benchmarked DL and non-DL based methods in generating SC-PET. Additionally, ARD-Net-generated SC-PET images displayed higher degree of concordance to SC-PET images in terms of radiomics features compared to non-DL and other DL approaches. Finally, the ensemble score suggested that ARD-Net exhibited significantly superior performance compared to benchmarked algorithms (p ≤ 0.01 post Bonferroni correction).

Conclusion: ARD-Net provides a robust framework to generate SC-PET from LC-PET images. ARD-Net generated SC-PET images exhibited superior performance compared other DL and non-DL approaches in terms of image-fidelity based metrics, task-based segmentation metrics, and minimal bias in terms of task-based quantification performance for preclinical PET imaging.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11423763PMC
http://dx.doi.org/10.1002/mp.17105DOI Listing

Publication Analysis

Top Keywords

sc-pet images
24
performance ard-net
12
compared benchmarked
12
performance
11
sc-pet
11
images
9
positron emission
8
emission tomography
8
photon-count levels
8
lc-pet images
8

Similar Publications

Background: Preclinical low-count positron emission tomography (LC-PET) imaging offers numerous advantages such as facilitating imaging logistics, enabling longitudinal studies of long- and short-lived isotopes as well as increasing scanner throughput. However, LC-PET is characterized by reduced photon-count levels resulting in low signal-to-noise ratio (SNR), segmentation difficulties, and quantification uncertainties.

Purpose: We developed and evaluated a novel deep-learning (DL) architecture-Attention based Residual-Dilated Net (ARD-Net)-to generate standard-count PET (SC-PET) images from LC-PET images.

View Article and Find Full Text PDF

Background: This study was conducted to clarify how patient body mass affects the image quality and quantification accuracy of images obtained using Zr PET/CT. Zr PET/CT images from time-of-flight (TOF) PET/CT and semiconductor (SC) PET/CT were obtained using three types (M, L, LL; corresponding to increasing patient body weight) of custom-made body phantoms designed similarly to the National Electrical Manufacturers Association (NEMA) IEC body phantom. The phantom data were analyzed visually and quantitatively to derive image quality metrics, namely detectability of the 10-mm-diameter hot sphere, percent contrast for the 10-mm-diameter hot sphere (Q), percent background variability (N), contrast-to-noise ratio (Q/N), and coefficient of variation of the background area (CV).

View Article and Find Full Text PDF

Scandium-44 has emerged as an attractive, novel PET radioisotope with ideal emission properties and half-life ( = 3.97 h, β = 632 keV) well matched to the pharmacokinetics of small molecules, peptides and small biologics. Conjugates of the current gold-standard chelator for Sc, 1,4,7,10-tetraaza-cyclododecane-1,4,7,10-tetraacetic acid (DOTA), require heating to achieve radiochemical complexation, limiting application of this isotope in conjunction with temperature-sensitive biologics.

View Article and Find Full Text PDF

AAZTA: An Ideal Chelating Agent for the Development of Sc PET Imaging Agents.

Angew Chem Int Ed Engl

February 2017

Bracco Imaging spa, Bracco Research Centre, Via Ribes 5, 10010, Colleretto Giacosa (TO), Italy.

Unprecedented fast and efficient complexation of Sc was demonstrated with the chelating agent AAZTA (AAZTA=1,4-bis(carboxymethyl)-6-[bis(carboxymethyl)]amino-6-methylperhydro-1,4-diazepine) under mild experimental conditions. The robustness of the Sc(AAZTA) chelate and conjugated biomolecules thereof is further shown by in vivo PET imaging in healthy and tumor mice models. The new results pave the way towards development of efficient Sc-based radiopharmaceuticals using the AAZTA chelator.

View Article and Find Full Text PDF

Evaluation of scatter limitation correction: a new method of correcting photopenic artifacts caused by patient motion during whole-body PET/CT imaging.

Nucl Med Commun

February 2016

aDepartment of Health Sciences, Division of Medical Quantum Science, Faculty of Medical Sciences, Kyushu University, Fukuoka bDepartment of Nuclear Medicine, Cancer Institute Hospital of Japanese Foundation for Cancer Research cResearch Team for Neuroimaging, Tokyo Metropolitan Institute of Gerontology, Tokyo dDepartment of Radiology, Chiba University Hospital, Chiba Prefecture, Japan.

Objective: Overcorrection of scatter caused by patient motion during whole-body PET/computed tomography (CT) imaging can induce the appearance of photopenic artifacts in the PET images. The present study aimed to quantify the accuracy of scatter limitation correction (SLC) for eliminating photopenic artifacts.

Methods: This study analyzed photopenic artifacts in (18)F-fluorodeoxyglucose ((18)F-FDG) PET/CT images acquired from 12 patients and from a National Electrical Manufacturers Association phantom with two peripheral plastic bottles that simulated the human body and arms, respectively.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!