The Westgard quality control (QC) rules are often applied in infectious diseases serology to validate the quality of results, but this requires a reasonable tradeoff between maximum sensitivity to errors and minimum false rejections. This article, in addition to illustrate the six sigma methodology in the QC management of the (anti-HCV Architect®) test, it discusses the main influencing factors on sigma value. Data from low positive and in-kit control materials spreading over 6 months and using four reagent kits, were used to calculate the precision of the test. The difference between the control material reactivity and the cut-off defined the error budget. Sigma values were > 6, which indicates that the method produces four erroneous results per million tests. The application of the six sigma concept made it possible to argue the choice of the new QC strategy (use of 13S rule with one positive control) and to relax the existing QC rules. This work provides a framework for infectious diseases serology laboratories to evaluate tests performances against a quality requirement and design an optimal QC strategy.