Introduction: Total joint replacement has become significantly more common as a treatment for advanced trapeziometacarpal joint osteoarthritis in recent years. The latest generation of prostheses with dual-mobility designs leads to very good functional results and low rates of loosening and dislocation in the short and medium term. Biomechanical studies showed that central placement and parallel alignment of the cup with respect to the proximal articular surface of the trapezium are crucial for both cup stability and prevention of dislocation. Despite correct positioning of the guidewire, incorrect placement or tilting of the inserted cup may occur, requiring immediate intraoperative revision.

Methods: The existing spherical and conical cup models in sizes 9 mm and 10 mm were transferred to a computer-aided design dataset. Depending on the intraoperative complication (tilting or incorrect placement), the revision options resulting from the various combinations of cup type and size were simulated and analyzed according to the resulting defect area and bony contact area.

Results: In well centered cups, a size 9 conical cup could be replaced by a size 9 spherical cup and still be fixed by press-fit. Conversely, a size 9 spherical cup could not be replaced by a size 9 conical cup, but only by a size 10 cup, of whatever shape. When a size 9 conical cup was tilted up to 20°, the best revision option was to resect the sclerotic margin and insert a size 10 conical cup deeper into the cancellous bone, to achieve the largest contact area with the surrounding bone. When a size 9 cup of whatever shape was poorly centered (misplaced with respect to the dorsopalmar or radioulnar line of the trapezium), placement should be corrected using a size 10 cup, combined with autologous bone grafting of the defect. Again, the size 10 conical cup showed the largest bony contact area.

Conclusion: Our computer-based measurements suggested options for intraoperative cup revision depending on cup shape and size and on type of misalignment with resulting bone defects. These suggestions, however, need to be confirmed in anatomic specimens before introducing them into clinical practice.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.hansur.2024.101712DOI Listing

Publication Analysis

Top Keywords

conical cup
24
size conical
20
cup
18
size
12
size cup
12
cup shape
12
incorrect placement
8
bony contact
8
cup replaced
8
replaced size
8

Similar Publications

Splash-cup plants disperse propagules via raindrops striking cup-shaped fruiting bodies. The seeds are ejected at velocities up to five times the impact speed of the raindrop and are dispersed up to 1 m from the parent plant. Here, we examine the effects of cup angles and the presence of seed mimics to understand the dynamics of this unique method of dispersal.

View Article and Find Full Text PDF

Introduction: Total joint replacement has become significantly more common as a treatment for advanced trapeziometacarpal joint osteoarthritis in recent years. The latest generation of prostheses with dual-mobility designs leads to very good functional results and low rates of loosening and dislocation in the short and medium term. Biomechanical studies showed that central placement and parallel alignment of the cup with respect to the proximal articular surface of the trapezium are crucial for both cup stability and prevention of dislocation.

View Article and Find Full Text PDF

There are scarce data regarding the rate of the occurrence of primary open-angle glaucoma (POAG) and visible lamina cribrosa pores (LCPs) in the eyes of individuals with African ancestry; the potential impact of these features on disease burden remains unknown. We recruited subjects with POAG to the Primary Open-Angle African American Glaucoma Genetics (POAAGG) study. Through regression models, we evaluated the association between the presence of LCPs and various phenotypic features.

View Article and Find Full Text PDF

Total Hip Arthroplasty has been a revolutionary technique in restoring mobility to patients with damaged hip joints. The introduction of modular components of the hip prosthesis allowed for bespoke solutions based on the requirements of the patient. The femoral stem is designed with a conical trunnion to allow for assembly of different femoral head sizes based on surgical requirements.

View Article and Find Full Text PDF

Two new species of Hypodontolaiminae, and , were isolated and described from the Yellow Sea, China. is characterized by four long cephalic setae, the amphidial fovea transverse oval in the male and slit-shaped in the female, the pharynx with a single posterior bulb, spicules curved and distally bifurcated, gubernaculum jointed, four (1+3) precloacal supplements papilliform, and the tail conical elongated with a short spinneret. is characterized by the buccal cavity with one large hollow dorsal tooth and two small subventral teeth, the pharynx with an obvious posterior bulb, spicules L-shaped and widened medially, gubernaculum boat-shaped, seven cup-shaped and equidistant precloacal supplements, and a long and gradually tapering tail.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!