Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Purpose: Standardization of x-ray cabinet irradiator dose, geometry, and calibration reporting is an ongoing process. Multi-tube designs have been introduced into the preclinical market and give a theoretical benefit but have not been widely assessed for use in preclinical irradiation conditions. The aim of this study was to report our experience commissioning a dual x-ray source cabinet irradiator (CIXD, Xstrahl Limited, United Kingdom) and assess the dose distribution for various experimental conditions.
Methods And Materials: Half-value layer (HVL) measurement, profile measurements, and output calibration were performed using a calibrated ion chamber. Constancy measurements were performed twice daily over 2 weeks to assess output fluctuations. Film measurements were completed using solid water to assess percent depth dose and homogeneity within the field and within variable thicknesses of solid water and phosphate-buffered saline solution. Film measurements were repeated for various arrangements of petri dishes filled with phosphate-buffered saline or water and in a 3D-printed mouse phantom.
Results: The x-ray tubes had a measured in-air output of 1.27 Gy/min. The HVL was 1.7 mm Cu. The upper and lower tubes both exhibited the heel effect, but when operated simultaneously, the effect was reduced. Ion chamber measurements revealed a 15% dose inhomogeneity within the tray area (18 × 18 cm). Film measurements in the petri dishes indicated minor nonuniformities in the arrangements of the experimental apparatus. Measurements from the mouse phantom with film agreed with ion chamber measurements for various phantom placements and orientations.
Conclusions: X-ray cell culture and animal irradiation with dual tube cabinet irradiation is efficient and robust when using established dosimetric tools to confirm output and homogeneity. The conditions assumed for calibrations are often not maintained during experiments. We have confirmed that inhomogeneities are present for single-tube use; however, they are reduced with simultaneous tube use. Additional dosimetric monitoring should be performed for each unique irradiation setup.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11063221 | PMC |
http://dx.doi.org/10.1016/j.adro.2024.101486 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!