  • The study analyzes real-world adverse events related to the use of the SynCardia total artificial heart (TAH) by reviewing FDA's MAUDE database from January 2012 to September 2020.
  • A total of 1,512 adverse events were reported, with infections (20.2%) and device malfunctions (20.1%) being the most frequent issues, while nearly 50% of cases recorded resulted in death.
  • The primary cause of death was multiorgan failure, highlighting the importance of early identification and management of complications post-TAH implantation for better patient outcomes.

Article Abstract

Background And Objectives: Real-world clinical data, outside of clinical trials and expert centers, on adverse events related to the use of SyncCardia total artificial heart (TAH) remain limited. We aim to analyze adverse events related to the use of SynCardia TAH reported to the Food and Drug Administration (FDA)'s Manufacturers and User Defined Experience (MAUDE) database.

Methods: We reviewed the FDA's MAUDE database for any adverse events involving the use of SynCardia TAH from 1/01/2012 to 9/30/2020. All the events were independently reviewed by three physicians.

Results: A total of 1,512 adverse events were identified in 453 "injury and death" reports in the MAUDE database. The most common adverse events reported were infection (20.2%) and device malfunction (20.1%). These were followed by bleeding events (16.5%), respiratory failure (10.1%), cerebrovascular accident (CVA)/other neurological dysfunction (8.7%), renal dysfunction (7.5%), hepatic dysfunction (2.2%), thromboembolic events (1.8%), pericardial effusion (1.8%), and hemolysis (1%). Death was reported in 49.4% of all the reported cases (n=224/453). The most common cause of death was multiorgan failure (n=73, 32.6%), followed by CVA/other non-specific neurological dysfunction (n=44, 19.7%), sepsis (n=24, 10.7%), withdrawal of support (n=20, 8.9%), device malfunction (n=11, 4.9%), bleeding (n=7, 3.1%), respiratory failure (n=7, 3.1%), gastrointestinal disorder (n=6, 2.7%), and cardiomyopathy (n=3, 1.3%).

Conclusions: Infection was the most common adverse event following the implantation of TAH. Most of the deaths reported were due to multiorgan failure. Early recognition and management of any possible adverse events after the TAH implantation are essential to improve the procedural outcome and patient survival.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11058438PMC
http://dx.doi.org/10.36628/ijhf.2023.0020DOI Listing

Publication Analysis

Top Keywords

adverse events
28
events
9
adverse
8
total artificial
8
artificial heart
8
food drug
8
drug administration
8
experience maude
8
syncardia tah
8
maude database
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!