Purpose: A systematic review and meta-analysis to evaluate the impact of extracorporeal carbon dioxide removal (ECCOR) on gas exchange and respiratory settings in critically ill adults with respiratory failure.

Methods: We conducted a comprehensive database search, including observational studies and randomized controlled trials (RCTs) from January 2000 to March 2022, targeting adult ICU patients undergoing ECCOR. Primary outcomes were changes in gas exchange and ventilator settings 24 h after ECCOR initiation, estimated as mean of differences, or proportions for adverse events (AEs); with subgroup analyses for disease indication and technology. Across RCTs, we assessed mortality, length of stay, ventilation days, and AEs as mean differences or odds ratios.

Results: A total of 49 studies encompassing 1672 patients were included. ECCOR was associated with a significant decrease in PaCO, plateau pressure, and tidal volume and an increase in pH across all patient groups, at an overall 19% adverse event rate. In ARDS and lung transplant patients, the PaO/FiO ratio increased significantly while ventilator settings were variable. "Higher extraction" systems reduced PaCO and respiratory rate more efficiently. The three available RCTs did not demonstrate an effect on mortality, but a significantly longer ICU and hospital stay associated with ECCOR.

Conclusions: ECCOR effectively reduces PaCO and acidosis allowing for less invasive ventilation. "Higher extraction" systems may be more efficient to achieve this goal. However, as RCTs have not shown a mortality benefit but increase AEs, ECCOR's effects on clinical outcome remain unclear. Future studies should target patient groups that may benefit from ECCOR. PROSPERO Registration No: CRD 42020154110 (on January 24, 2021).

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11061932PMC
http://dx.doi.org/10.1186/s13054-024-04927-xDOI Listing

Publication Analysis

Top Keywords

gas exchange
12
ventilator settings
12
exchange ventilator
8
systematic review
8
review meta-analysis
8
patient groups
8
"higher extraction"
8
extraction" systems
8
eccor
6
effects extracorporeal
4

Similar Publications

To define training zones, ventilatory thresholds (VTs) are commonly established by cardiopulmonary gas-exchange analysis during incremental exercise tests. Portable near-infrared spectroscopy (NIRS) devices have emerged as a potential tool for detecting these thresholds by monitoring muscle oxygenation. This study evaluated the accuracy of NIRS measurements to determine VTs or critical power (CP) based on muscle oxygen saturation and assesses the device's consistency across 2 constant-load tests.

View Article and Find Full Text PDF

[Acute respiratory distress syndrome : Pathophysiology, definition and treatment strategies].

Med Klin Intensivmed Notfmed

January 2025

Universitätsklinik für Innere Medizin I, Medizinische Universität Wien, Allgemeines Krankenhaus der Stadt Wien, Währinger Gürtel 18-20, 1090, Wien, Österreich.

Acute respiratory distress syndrome (ARDS) is defined as an acute inflammatory syndrome leading to increased pulmonary capillary leakage and subsequent interstitial and alveolar pulmonary edema. Hypoxia is the predominant symptom. The definition of ARDS comprises acute onset, bilateral patchy infiltration on chest X‑ray and a reduction of the ratio of arterial partial pressure of oxygen (PaO) to the fraction of inspired oxygen (FiO), which also determines the classification into mild (≤ 300), moderate (≤ 200) and severe (≤ 100) ARDS.

View Article and Find Full Text PDF

Neutrophil extracellular traps (NETs) were detected in blood samples and in cellular deposits of oxygenator membranes during extracorporeal membrane oxygenation (ECMO) therapy and may be responsible for thrombogenesis. The aim was to evaluate the effect of the base material of gas fiber (GF, polymethylpentene) and heat exchange (HE) membranes and different antithrombogenic coatings on isolated granulocytes from healthy volunteers under static culture conditions. Contact of granulocytes with membranes from different ECMO oxygenators (with different surface coatings) and uncoated-GFs allowed detection of adherent cells and NETotic nuclear structures (normal, swollen, ruptured) using nuclear staining.

View Article and Find Full Text PDF

Does a priming warm-up influence the incidence of [Formula: see text] during a ramp test and verification phase?

PLoS One

January 2025

Center for Research and Education in Special Environments, Department of Exercise and Nutrition Sciences, University at Buffalo, Buffalo, NY, United States of America.

Objective: This study compared the effects of two different warm-up protocols (normal vs. priming) on the oxygen plateau ([Formula: see text]) incidence rate during a ramp test. It also compared the cardiopulmonary responses during the ramp test and subsequent verification phase.

View Article and Find Full Text PDF

Thermoelectric Energy Harvesting for Exhaust Waste Heat Recovery: A System Design.

ACS Appl Mater Interfaces

January 2025

Department of Materials Science and Engineering, Pennsylvania State University, University Park, Pennsylvania 16802, United States.

Thermal energy harvesting for high-speed moving objects is particularly promising in providing an efficient and sustainable energy source to enhance operational capabilities and endurance. Thermoelectric (TE) technology, by exploiting temperature gradients between a heat source and ambient temperature, can provide a continuous power supply to such systems, reducing the reliance on conventional batteries and extending operation times. However, the integrated thermoelectric generator (TEG) system design research is far behind materials development.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!