Agreement of Serial iCare HOME2 and Goldmann Applanation Tonometry.

Ophthalmol Glaucoma

Department of Ophthalmology, Palo Alto Medical Foundation, Palo Alto, California; Byers Eye Institute, Horngren Family Vitreoretinal Center, Department of Ophthalmology, Stanford University School of Medicine, Palo Alto, California.

Published: September 2024

Purpose: To assess agreement of iCare HOME2 and Goldmann applanation tonometry over a wide range of intraocular pressure (IOP).

Design: A prospective, observational cohort study.

Subjects: Twenty-six adult patients undergoing intravitreal injection, which temporarily raises IOP, were recruited from the Palo Alto Medical Foundation Retina Clinic between October 2022 and February 2023.

Methods: Subjects had serial iCare HOME2 (IOP) and Goldmann applanation (IOP) IOP measurements before and at 0 and 5 to 10 minutes after injection. Baseline IOPs and pachymetry were taken in both eyes.

Main Outcome Measures: Correlation between IOP and IOP was tested by within-subjects intraclass correlation coefficient (ICC) for repeated measures. Agreement between IOP and IOP was evaluated by a Bland-Altman plot with correction for multiple measurements. The difference between IOP and IOP was evaluated between eyes at baseline (Pearson's r) and within the injected eye over different timepoints (ICC for absolute agreement). Linear regression was used to evaluate the effects of age, sex, glaucoma, and corneal thickness.

Results: The mean IOP and IOP were 25.3 (range: 9-55) and 23.5 (range: 8-56) mmHg, respectively. The correlation between IOP and IOP was 0.99 (P < 0.001). The mean difference (IOP - IOP) was 2.2 mmHg (95% limits of agreement: -3.4 to 7.8 mmHg). The bias in measurements was correlated between eyes (r, 0.68; P < 0.001) and in the injected eye across all timepoints (ICC, 0.86; 95% CI, 0.75-0.93), but did not show a relationship with age, sex, glaucoma, or corneal thickness.

Conclusions: IOP and IOP showed excellent correlation; however, there was a stable bias toward IOP being higher than IOP over a large range of IOP.

Financial Disclosure(s): Proprietary or commercial disclosure may be found in the Footnotes and Disclosures at the end of this article.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ogla.2024.04.007DOI Listing

Publication Analysis

Top Keywords

iop iop
28
iop
19
icare home2
12
goldmann applanation
12
serial icare
8
home2 goldmann
8
applanation tonometry
8
correlation iop
8
iop evaluated
8
injected eye
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!