Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: In vitro assessment is mandatory for artificial heart valve development. This study aims to investigate the effects of pulse duplicator features on valve responsiveness, conduct a sensitivity analysis across valve prosthesis types, and contribute on the development of versatile pulse duplicator systems able to perform reliable prosthetic aortic valve assessment under physiologic hemodynamic conditions.
Methods: A reference pulse duplicator was established based on literature. Further optimization process led to new designs that underwent a parametric study, also involving different aortic valve prostheses. These designs were evaluated on criteria such as mean pressure differential and pulse pressure (assessed from high-fidelity pressure measurements), valve opening and closing behavior, flow, and regurgitation. Finally, the resulting optimized setup was tested under five different hemodynamic settings simulating a range of physiologic and pathologic conditions.
Results: The results show that both, pulse duplicator design and valve type significantly influence aortic and ventricular pressure, flow, and valve kinematic response. The optimal design comprised key features such as a compliance chamber and restrictor for diastolic pressure maintenance and narrow pulse pressure. Additionally, an atrial reservoir was included to prevent atrial-aortic interference, and a bioprosthetic valve was used in mitral position to avoid delayed valve closing effects.
Conclusion: This study showed that individual pulse duplicator features can have a significant effect on valve's responsiveness. The optimized versatile pulse duplicator replicated physiologic and pathologic aortic valve hemodynamic conditions, serving as a reliable characterization tool for assessing and optimizing aortic valve performance.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/aor.14757 | DOI Listing |