Do Balloon Catheters have a Different Radial Force Along Their Longitudinal Axis?

Cardiovasc Intervent Radiol

Department of Radiology, Ajou University School of Medicine, Ajou University Hospital, Suwon, Republic of Korea.

Published: September 2024

Purpose: This experimental study was designed to compare radial forces between the central portion and both ends of balloon catheters when dilating stenosis.

Materials And Methods: Three balloon catheters of 6 and 8 mm in diameter and of variable length were tested: Mustang, Conquest, and Genoss PTA. Cylindrical modules to position balloon catheters and install the measuring tip during radial force measurements were made using a 3D printer. The measuring tip created 20% stenosis at the inner lumen. Both ends and center of the balloon catheter were located at the measuring tip. The radial force was measured after inflating the balloon catheter to the rated burst pressure.

Results: For the different diameters and lengths of balloon catheters and cylinder sizes, the median inccenter, the radial rease in radial force at the distal end compared to the center was 16.5% (range: 9.8-35.2%) for Mustang, 12.4% (range: 10.3-25.5%) for Genoss, and 7.4% (range: -0.3-13.1%) for Conquest balloon catheters. Similarly, compared to that at the force at the proximal end was 10.8% greater (range: -2.9-18.3%) for Mustang, 9.9% greater (range: 3.9-22.3%) for Genoss, and 7.3% greater (range: -1.3-12.4%) for Conquest catheters.

Conclusion: The radial force is greater at both ends of the balloon than at the central portion, especially at the distal end. Dilation using the distal end of the balloon catheter is a practical method that can be applied in clinical practice without additional devices when encountering resistant stenosis, especially with semi-compliant balloons.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11379730PMC
http://dx.doi.org/10.1007/s00270-024-03716-xDOI Listing

Publication Analysis

Top Keywords

balloon catheters
24
radial force
20
balloon catheter
12
greater range
12
balloon
10
central portion
8
ends balloon
8
measuring radial
8
radial
7
force
6

Similar Publications

Management of Anterior Choroidal Artery Aneurysms: A Retrospective Cohort Study.

Brain Sci

December 2024

Division of Neurointerventional Radiology, UMass-Chan Lahey Department of Radiology, Lahey Hospital and Medical Centre-Beth Israel Lahey Health, Burlington, MA 01805, USA.

Anterior choroidal artery (AChoA) aneurysms pose a challenge for both endovascular and clipping procedures. The eloquent territory supplied by the parent vessel has limited collateralization and its compromise can lead to significant morbidity. This study aims to analyze the clinical outcomes and procedure-related complications of clipping and endovascular treatment of AChoA aneurysms to aid physician decision making.

View Article and Find Full Text PDF

Background: Assessing maternal pain and satisfaction following administration of paracetamol vs. placebo prior to catheter balloon placement.

Methods: Primiparous women at term admitted for medically-indicated labor induction were randomized to receive intravenous paracetamol 1 gram in 100cc normal saline (N=71) or placebo of 100cc normal saline (N=70) prior to catheter balloon insertion.

View Article and Find Full Text PDF

Intrauterine Adhesions (IUAs) are characterized by endometrial damage due to endometritis or curettage. Currently, the gold standard for IUA treatment is hysteroscopy, which enables the dissolution of IUA through mechanical or electrosurgical energy. Common strategies to prevent recurrence include the insertion of a balloon catheter or IUD in the uterus.

View Article and Find Full Text PDF

Proximal protection devices for carotid artery stenting - A benchtop assessment of flow reversal performance.

AJNR Am J Neuroradiol

January 2025

From the Department of Radiology (J.L., E.A.B., C.B., J.C., R.K., W.B., D.F.K), and Department of Neurologic Surgery (Y.C.S., R.K., W.B.), Mayo Clinic, Rochester, MN, United States; Department of Stroke Research (J.L.), Vall d'Hebron Research Institute, Barcelona, Spain; From the Global Institute of Future Technology (Y.L.), Shanghai Jiao Tong University, Shanghai, China; Department of Neurointerventional Radiology (J.C.), Bicetre University Hospital, Le Kremlin Bicetre, France.

Background And Purpose: Proximal protection devices, such as TransCarotid Artery Revascularization (TCAR, SilkRoad Medical, Sunnyvale), aim to yield better outcomes in carotid artery stenting (CAS) than distal protection devices by preventing plaque embolization to the brain. However, transfemoral catheters may not fully reverse flow from the external carotid artery (ECA) to the internal carotid artery (ICA). We assess a new balloon-sheath device, Femoral Flow Reversal Access for Carotid Artery Stenting (FFRACAS), for this purpose.

View Article and Find Full Text PDF

Drug-Coated Balloon-Based Treatment of Left Main True Bifurcation Lesion.

Catheter Cardiovasc Interv

January 2025

Department of Internal Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Gangnam-gu, South Korea.

Background: There is a paucity of data regarding drug-coated balloon (DCB) treatment in the context of left main (LM) true bifurcation lesions.

Aims: The aim of this study was to evaluate the safety and efficacy of DCB-based treatment for unprotected LM true bifurcation lesions.

Methods: A total of 39 patients with LM true bifurcation lesion (Medina: 1,1,1/0,1,1/1,0,1) who were successfully treated with DCB alone or in combination with drug-eluting stent (DES) were retrospectively enrolled into the DCB-based group.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!