Using nonlinear auto-regressive with exogenous input neural network (NNARX) in blood glucose prediction.

Bioelectron Med

Automatic control department, Tabin Institute For Metallurgical Studies, Iron and Steel street, Tabbin, Helwan, Cairo, Egypt.

Published: April 2024

Background: Predicting of future blood glucose (BG) concentration is important for diabetes control. Many automatic BG monitoring or controlling systems use BG predictors. The accuracy of the prediction for long prediction time is a major factor affecting the performance of the control system. The predicted BG can be used for glycemia management in the form of early hypoglycemic/hyperglycemic alarms or adjusting insulin injections. Recent developments in continuous glucose monitoring (CGM) devices open new opportunities for glycemia management of diabetic patients. Many of those systems need prediction for long prediction horizons to avoid going through hypo or hyperglycemia.

Methods: In this article a nonlinear autoregressive exogenous input neural network (NNARX) is proposed to predict the glucose concentration for longer prediction horizons (PHs) than that was obtained previously with an established recurrent neural network (RNN). The proposed NNARX is a modified version from our previously published RNN with different initialization and building technique but has the same architecture. The modification is based on starting with building nonlinear autoregressive exogenous input model using MATLAB and train it, then close the loop to get NNARX network.

Results: The results of using the proposed NNARX indicate that the proposed NNARX is better in prediction and stability than unmodified RNN as PH becomes higher than 45 minutes.

Conclusions: Modification in RNN building extends the ability of the prediction till 100 minutes. It performs statistically significant improvements in the FIT and RMSE values for 100 minutes prediction. It also decreases root mean squared error (RMSE) for both 45 and 60 minutes of prediction.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11636905PMC
http://dx.doi.org/10.1186/s42234-024-00141-wDOI Listing

Publication Analysis

Top Keywords

exogenous input
12
neural network
12
proposed nnarx
12
prediction
10
input neural
8
network nnarx
8
blood glucose
8
glucose concentration
8
prediction long
8
long prediction
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!