Clinical Relevance: Tear meniscus height (TMH) is an important clinical marker in dry eye diagnosis and management.

Purpose: To evaluate the reproducibility and agreement of TMH measurements in non-clinical participants using the Oculus Keratograph 5 M, Medmont Meridia, and Spectral-domain optical coherence tomography (Spectralis SD-OCT).

Methods: Fifty-six participants (mean 43.8 ± 22.4 years) were recruited for this cross-sectional study. Image acquisitions were performed on the three devices, sequentially and randomized. The repeatability and reproducibility of inter-observer and inter-device analysis were performed. Repeated measures ANOVA and Bland-Altman Plots were used to evaluate the agreement between devices.

Results: The mean TMH with the Oculus Keratograph 5 M, Medmont Meridia and Spectralis SD-OCT were 0.29 ± 0.16 mm, 0.24 ± 0.09 mm and 0.27 ± 0.16 mm, respectively. There were no significant inter-observer differences (paired t-tests,  < 0.001). All the devices exhibited good inter-observer reliability (ICC ≥ 0.877), and good repeatability (CV ≤ 16.53%). Inter-device reliability is moderate (ICC = 0.621,  < 0.001). Repeated measures ANOVA revealed that TMH measurements given by the Spectralis SD-OCT are not significantly different from the Oculus Keratograph 5 M ( = 0.19) and the Medmont Meridia ( = 0.38). TMH measurements from Oculus Keratograph 5 M were significantly higher than those from Medmont Meridia ( = 0.02). Correlations between the mean TMH and the difference in the TMH measurements were positive for Oculus Keratograph 5 M and Medmont Meridia (r = 0.62,  < 0.001), negative for Medmont Meridia and Spectralis SD-OCT (r = -0.59,  < 0.001), and not significant for Oculus Keratograph 5 M and Spectralis SD-OCT (r = 0.05,  = 0.74). A strong correlation was found for TMH measured with all devices (r = 0.55 to 0.81,  < 0.001).

Conclusions: The Oculus Keratograph 5 M, Medmont Meridia, and Spectralis SD-OCT provide reliable and reproducible inter-observer TMH measurements. Inter-device reliability is moderate, with a close correlation between Spectralis SD-OCT and the Oculus Keratograph 5 M. Oculus Keratograph 5 M and Medmont Meridia are repeatable devices appropriate for the measurement of TMH, but they are not interchangeable in clinical practice.

Download full-text PDF

Source
http://dx.doi.org/10.1080/08164622.2024.2341833DOI Listing

Publication Analysis

Top Keywords

tear meniscus
8
meniscus height
8
spectral-domain optical
8
optical coherence
8
coherence tomography
8
oculus keratograph
8
keratograph 5 m
8
5 m medmont
8
medmont meridia
8
height agreement
4

Similar Publications

Purpose: To evaluate the efficacy and safety of intense pulsed light (IPL) combined with meibomian gland expression (MGX) for the treatment of dry eye disease and meibomian gland dysfunction associated with chronic Stevens-Johnson syndrome and toxic epidermal necrolysis.

Methods: This prospective noncomparative interventional study included 29 patients (58 eyes) who underwent 3 sessions of IPL and MGX at 2-week intervals. Subjective symptoms (ocular surface disease index score) and objective dry eye tests: matrix metalloproteinase 9, tear meniscus height, bulbar redness score, tear film lipid layer thickness (LLT), Schirmer I test, conjunctival and corneal staining, meibomian gland loss, MGX score [meibomian gland score (MGS)], and tear break-up time were assessed at the baseline and after 4, 8, and 12 weeks.

View Article and Find Full Text PDF

The meniscus plays a vital role in knee biomechanics, particularly in load distribution and stability. Meniscus root tears (MRTs) compromise these functions, resulting in biomechanical alterations and knee osteoarthritis. The effectiveness of different MRT treatments is not yet well defined.

View Article and Find Full Text PDF

Background: Medial meniscus posterior root tears (MMPRTs) significantly contribute to knee dysfunction, leading to abnormal biomechanics and accelerated cartilage degeneration. Arthroscopic transtibial pullout and all-inside repair are two commonly used techniques for treating MMPRTs, each with unique advantages and limitations.

Objective: To compare the clinical and functional outcomes of the transtibial pullout and all-inside repair techniques in the treatment of MMPRTs, with a focus on postoperative recovery, knee function, and complications.

View Article and Find Full Text PDF

Introduction: Despite the good clinical outcomes of osteochondral autograft transplantation (OAT), reports of re-arthroscopic findings after OAT have been limited to short-term, and there are no reports of findings after long-term follow-up. This is the first report that describes re-arthroscopic findings long-term after OAT.

Case Report: A male patient underwent OAT on the lateral femoral condyle (LFC) of the knee and lateral meniscus (LM) repair at the age of 45.

View Article and Find Full Text PDF

Purpose: To compare the meibographies and dry eye parameters of paretic vs non-paretic sides of patients with a facial palsy diagnosis.

Patients And Methods: Twenty patients with unilateral facial palsy were recruited and the severity of the disease was staged using the House-Brackmann scale. A comprehensive dry eye evaluation was performed using the Oculus 5M Keratograph.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!