Background: The purpose of this study is to evaluate the performance of recently developed tumor marker clinical kits in China, with the aim of encouraging local medical technology innovation and thus narrowing the research and development gap with foreign kits.
Methods: The newly established reagent kits were analyzed on the TESMI F3999-Luminex200 flow lattice instrument to verify precision, sensitivity (blank limit), linearity, anti-interference ability, carry-over contamination rate, hook effect, and reference interval verification. Additionally, the newly established reagent kits were compared to other commercially available detection kits (reference reagent kits) to analyze the correlation between the two types of kits.
Results: The intra-assay and inter-assay precision had coefficients of variations (CVs) less than 3.50% and 6.91%, respectively. The tumor marker blank limits were lower than the manufacturer's statement. The newly established reagent kits demonstrated excellent linearity (r > 0.99). Rheumatoid factor, triglycerides, bilirubin, and hemoglobin did not have significant interference with the determination of tumor markers. The carry-over contamination rates were all much lower than 3%. At extremely high concentrations of AFP (277,335 ng/mL and 1,031,424 ng/mL), the measured tumor marker values were higher than the upper limit of the linear range and no hook effect occurred. The reference interval was suitable for use in clinical laboratory settings. Correlation analysis indicated a satisfactory relevance and consistency between the newly developed reagent kits and reference reagent kits, with correlation coefficients of r > 0.967 among 654 patients and healthy individuals.
Conclusions: The newly developed reagent kits for tumor markers performed well in all evaluated parameters, having the potential for clinical promotion and application.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.7754/Clin.Lab.2023.230813 | DOI Listing |
J Transl Med
January 2025
Tumour Biology and Immunology Laboratory, Research Branch, Sidra Medicine, Doha, Qatar.
Background: FFPE tissue samples are commonly used in biomedical research and are a valuable source for next-generation sequencing in oncology, however, extracting RNA from these samples can be difficult the quantity and quality achieved can impact the downstream analysis. This study compared the effectiveness of seven different commercially available RNA extraction kits specifically designed for use with FFPE samples in terms of the quantity and quality of RNA recovered.
Methods: This study used 9 samples of FFPE tissue from three different types of tissue (Tonsil, Appendix and lymph node of B-cell lymphoma) to evaluate RNA extraction methods.
AIDS Res Ther
January 2025
Department of Mathematics, Physics and Informatics, Mkwawa University College of Education, Iringa, Tanzania.
Introduction: The introduction of the HIV self-test kit in the early 2000s was a major breakthrough in combating HIV. This study determines the social demographic and sexual behaviour driving the use of HIV self-test kits.
Method: The study used secondary data obtained from Tanzania DHS-MIS 2022.
J Vis Exp
December 2024
Department of Epigenetics and Molecular Carcinogenesis, University of Texas MD Anderson Cancer Center; Department of Gynecologic Oncology and Reproductive Medicine, University of Texas MD Anderson Cancer Center;
The CUT&RUN technique facilitates detection of protein-DNA interactions across the genome. Typical applications of CUT&RUN include profiling changes in histone tail modifications or mapping transcription factor chromatin occupancy. Widespread adoption of CUT&RUN is driven, in part, by technical advantages over conventional ChIP-seq that include lower cell input requirements, lower sequencing depth requirements, and increased sensitivity with reduced background signal due to a lack of cross-linking agents that otherwise mask antibody epitopes.
View Article and Find Full Text PDFCardiol Young
January 2025
Department of Cardiovascular, The Second Hospital of Hebei Medical University, Shijiazhuang, China.
Mikrobiyol Bul
October 2024
University of Health Sciences, Ankara Bilkent City Health Application and Research Center, Clinic of Medical Microbiology, Ankara, Türkiye.
The severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2) virus has mutated at a high rate since the beginning of the pandemic, leading to the formation of different variants. Alpha, Beta, Gamma, Delta and Omicron have emerged as concerning variants identified by the World Health Organization (WHO). The Omicron variant and its sublineages became dominant worldwide in 2022.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!