Introduction: The Dual Shell Encompass Fibreplast™ System (DS-Encompass) by CQ Medical™ is validated for frameless immobilization in stereotactic brain radiotherapy. An alternative mask model has been proposed with the rear shell replaced by a Moldcare® cushion (M-Encompass). To validate the use of this model in our cranial stereotactic workflow method including HyperArc™, we performed a prospective randomized study comparing inter-and intrafractional motion and patients comfort between both masks.

Materials & Methods: A prospective randomized study between DS-Encompass and M-Encompass was conducted involving 60 participants. Stratification between DS-Encompass and M-Encompass was carried out based on the fractionation scheme. Treatment plans were created with HyperArc™. During treatment, surface guidance was used for patient positioning and monitoring. A pre-treatment cone-beam CT (CBCT) was acquired to correct interfractional motion and a post-treatment CBCT was acquired to quantify the intrafractional motion. Patients reported comfort was analyzed with a Likert-scale at the end of the treatment. Unpaired t-tests were conducted to determine the level of significance.

Results: No significant difference in interfractional translations is present. A significant difference is revealed in roll-axis rotation, where DS-Encompass allows for smaller deviations. Since interfractional motion can be corrected through daily CBCT-scans and 6D-couch corrections, they are clinically irrelevant. Intrafractional motion does not differ significantly and remains below 0.5 mm and 0.5° for both systems. There is no statistical difference in patient-reported comfort.

Conclusion: We conclude that Encompass with Moldcare offers a safe alternative to Duall Shell Encompass for non-coplanar stereotactic brain radiotherapy. There is no significant difference in intrafractional motion nor difference in comfort levels.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11017338PMC
http://dx.doi.org/10.1016/j.tipsro.2024.100249DOI Listing

Publication Analysis

Top Keywords

intrafractional motion
16
prospective randomized
12
randomized study
12
study comparing
8
frameless immobilization
8
cranial stereotactic
8
shell encompass
8
stereotactic brain
8
brain radiotherapy
8
motion patients
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!