The Holmium:YAG laser has been the gold standard for laser lithotripsy over the past three decades and, since the late 1990s, also for prostate enucleation. Pulsed thulium fibre lasers (TFL) demonstrated their efficacy in in-vitro experiments and were introduced to the market a few years ago. Initial clinical results for TFL in lithotripsy and enucleation are very promising. In addition to TFL, a pulsed Thulium:YAG solid-state laser has been introduced, but clinical data for this laser are currently limited. This article aims to review the key technological differences between Ho:YAG lasers and pulsed thulium lasers and compare/discuss the initial clinical results for stone lithotripsy and laser enucleation.In-vitro studies have demonstrated the technical superiority of TFL compared with Ho:YAG lasers. However, as TFL is still a new technology, only limited studies are available to date, and optimal settings for lithotripsy have not been established. For enucleation, the differences of TFL compared with a high-power Ho:YAG laser seem to be clinically irrelevant. Initial studies on pulsed Tm:YAG lasers show good results, but there continues to be a lack of comparative studies.Based on the current literature, pulsed thulium lasers have the potential of being an alternative to Ho:YAG lasers. However, further studies are necessary to determine the optimal laser technology for enucleation and lithotripsy of urinary stones, considering all parameters, including efficacy, safety, and cost.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1055/a-2286-1379 | DOI Listing |
Urolithiasis
December 2024
Groupe de Recherche Clinique sur la Lithiase Urinaire, GRC n°20, Hôpital Tenon, Sorbonne Université, Paris, 75020, France.
To assess the accuracy of Kidney Stone Calculator(KSC), a software designed for surgical planning, in predicting the duration of lithotripsy during flexible ureteroscopy(FURS) when using the novel pulsed-Thulium: YAG(p-Tm: YAG) laser. From February to August 2023, a single-center prospective study was conducted, including patients with kidney or ureteral stones through non-contrast computed tomography(NCCT), who underwent FURS with p-Tm: YAG laser lithotripsy. KSC used three-dimensional segmentation of the stones from NCCT images, along with an interactive user interface for laser settings, to estimate the stone volume(SV) and the lithotripsy duration(LD).
View Article and Find Full Text PDFPurpose: Percutaneous nephrolithotomy (PNL) is a controlled grade IV renal trauma and intraoperative bleeding is the most crucial complication. Contemporarily, pulsed mode thulium laser has been reported to confer adequate hemostasis. Therefore, the primary outcome of this study was to assess safety and efficacy of thulium laser-based hemostasis during PNL.
View Article and Find Full Text PDFWorld J Urol
December 2024
Department of Urology, University of Freiburg Medical Center, Freiburg, Germany.
Objective: To evaluate the impact of three different AEEP techniques on the training performance of novices using a realistic hydrogel prostate phantom model.
Material And Methods: The experimental setup utilized realistic prostate phantom model provided by the Max Planck Institute for Intelligent Systems, Germany. For the enucleation, we utilized a new solid-state pulsed thulium laser (Thulio®, Dornier MedTech, Weßling, Germany).
Lasers Med Sci
December 2024
Department of Biotechnology, Mizoram University, Aizawl, Mizoram, India.
Simulation studies on temperature distribution in laser ablation help predict ablation rates, laser settings, and thermal damage. Despite the limited number of reported numerical studies on the temperature distribution of kidney fluid, there is no simulation study for kidney stone temperature distribution. We employ a numerical approach to study the kidney stone temperature distribution and predict ablation rates, which is an important parameter for clinical lithotripsy.
View Article and Find Full Text PDFLasers Med Sci
December 2024
Groupe de Recherche Clinique sur la Lithiase Urinaire, GRC n°20, Hôpital Tenon, Sorbonne Université, Paris, 75020, France.
To compare the pulsed-Thulium: YAG(p-Tm: YAG) and Thulium Fiber(TFL) lasers in terms of efficiency and safety profiles during flexible ureteroscopy(fURS) and endocorporeal laser lithotripsy(ELL). A prospective single-center open-label comparative study included consecutive patients with ureteral and renal stones who underwent fURS using Thulio(p-Tm: YAG, Dornier©,Germany) or TFL Drive(TFL, Coloplast©,Danemark), with 270 μm and 150/200μm laser fibers(LF), respectively. fURS were performed by a single operator in each group.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!