Purpose: To evaluate clinical outcomes after bilateral or contralateral implantation of the Gemetric™ (G) and Gemetric™ Plus (GPlus) diffractive trifocal intraocular lenses (IOLs).

Methods: This was a prospective, randomized, multi-center open-label study comparing clinical results and subjective patient responses around 6 months after implantation of the study lenses (toric and non-toric) in three different groups (bilateral G, bilateral GPlus and contralateral G/GPlus implantation). Results included the manifest refraction, uncorrected and distance corrected monocular and binocular visual acuity (VA) at distance, intermediate and near; the defocus curve; contrast sensitivity; and patient reported outcomes regarding spectacle independence, satisfaction and visual disturbances.

Results: There was no statistically significant difference in the mean refraction spherical equivalent between the two lens models (p = 0.33) or between the toric and non-toric lenses (p = 0.06). Monocular VA was better at distance with the G lens and better at near with the GPlus lens (p < 0.01). Mean binocular VA was better than 0.1 logMAR at all distances for all groups, both uncorrected and distance corrected. The mean binocular distance corrected VA was better than 0.15 logMAR from 0.0 D to -3.50 D for all groups. All VA data for the contralateral group was as good or better than for the bilateral GPlus group. Questionnaire results showed no difference between groups for the frequency, severity, or degree of bother of visual disturbances (p > 0.24).

Conclusion: The two diffractive trifocal IOLs studied here may be used either bilaterally or contralaterally for the correction of presbyopia in cataract patients, providing excellent visual acuity with low levels of visual disturbances and high rates of overall spectacle independence. Bilateral Gemetric implantation resulted in slightly better distance and intermediate vision while contralateral implantation provided slightly better near vision. There was no apparent advantage to implanting the GPlus IOL bilaterally.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10999202PMC
http://dx.doi.org/10.2147/OPTH.S456007DOI Listing

Publication Analysis

Top Keywords

diffractive trifocal
12
distance corrected
12
clinical outcomes
8
trifocal iols
8
contralateral implantation
8
toric non-toric
8
bilateral gplus
8
uncorrected distance
8
visual acuity
8
distance intermediate
8

Similar Publications

An intraocular lens (IOL) replaces the natural crystalline lens during cataract surgery, and although the vast majority of implants have simple optics, "advanced technology" IOLs have multifocal and extended depth of focus (EDOF) properties. Optical concepts are evaluated here, with image contrast, focal range, and unwanted visual phenomena being the primary concerns. Visual phenomena with earlier bifocal diffractive lenses led to alternative diffractive designs (trifocals, etc.

View Article and Find Full Text PDF

Visual Performance Correlation with Corneal Aberrometric Profile and Pupil Size After Implantation of a Trifocal Hydrophobic IOL.

Clin Ophthalmol

December 2024

Département de Recherche Clinique, Institut Ophtalmologique de l'Ouest (IOO) Jules Verne, Nantes, France.

Purpose: To evaluate clinical outcomes obtained after cataract surgery involving the implantation of a trifocal hydrophobic intraocular lens (IOL) and to determine if pupil size and the corneal aberrometric profile correlate to visual acuity at different distances.

Methods: 49 patients (98 eyes) underwent bilateral cataract surgery with the placement of FineVision HP IOLs for presbyopia and were assessed at 1- and 3- to 6-months post-surgery. Postoperatively, refraction, monocular and binocular uncorrected and corrected distance visual acuity (UDVA and CDVA), uncorrected intermediate visual acuity (UIVA), uncorrected near visual acuity (UNVA) and the binocular defocus curve were measured.

View Article and Find Full Text PDF

Introduction: Diffractive trifocal intraocular lenses (IOLs) provide good vision at distance, intermediate, and near, but can also cause positive dysphotopsias. This meta-analysis pooled published evidence on visual disturbances after bilateral implantation of the PanOptix (TFNTXX) IOL for patients undergoing cataract surgery.

Method: A systematic literature search was conducted in PubMed and congress presentations from April 2021 to December 2022 to identify studies with patient-reported outcomes on the incidence of visual disturbances (starbursts, halos, glare) post bilateral implantation of PanOptix IOL during cataract surgery.

View Article and Find Full Text PDF

This study investigates whether a diffractive presbyopia-correcting multifocal intraocular lens disrupts the favorable interaction between chromatic and monochromatic aberrations in the eye. This is analyzed not only for distant objects but also for closer viewing distances, where the lens utilizes different diffraction orders depending on its design. We consider diffractive designs based on the zero-diffraction order for far vision and the first diffraction order for near vision (i.

View Article and Find Full Text PDF

Visually relevant on-bench through-focus analysis of intraocular lenses.

Biomed Opt Express

December 2024

Laboratorio de Óptica, Universidad de Murcia, Campus de Espinardo, 30100 Murcia, Spain.

Cataract surgery involves the implantation of an intraocular lens (IOL) to replace the opacified crystalline lens. Monofocal IOLs, the most common type, are intended to have the eye in focus at a given distance, usually at infinity. Simultaneous vision IOLs (SVIOLs) and extended depth of focus (EDOF) aim to minimize postoperative dependence on spectacles by providing either multiple foci or an extended depth of focus.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!