Purpose: Computed tomography (CT)-based robotic system for total knee arthroplasty (TKA) has shown improved accuracy compared to conventional. This study was designed to (1) confirm the accuracy of the robotic system in achieving the plan and (2) establish the alignment and positioning deviation between final components and planning, by measuring the discrepancy between final implant alignment and the corresponding planned cut.

Methods: Ninety-six cementless robotic-arm assisted (RA) TKAs were assessed. Bone resections were performed using the haptically controlled robotic arm. Alignment in the coronal and sagittal plane and resection depth of the distal femoral and proximal tibial cuts were recorded with a navigation planar probe. After final components were impacted, the probe was positioned on each implant surface to determine its alignment and positioning.

Results: The mean tibial resections and implanted tibial component's positioning were 0.4 mm (standard deviation, SD: 0.6) and 0.9 mm (SD: 0.8), respectively, higher than planned (p < 0.01). The tibial sagittal cut had 19/96 cases (19.8%) of ±1° outliers from plan. In 40/96 cases (41.7%), the tibial component was more prominent than planned of more than 1 mm. The mean femoral resections and impacted femoral component's positioning was 0.1 mm (SD: 0.8) and 0.2 mm (SD: 0.7), respectively, higher than planned. In 23/96 cases (24.0%), the femoral sagittal cut and femoral component coronal alignment deviated more than ±1° from plan.

Conclusions: The computed tomography-based robotic-assisted TKA system showed good accuracy regarding bone preparation and component's positioning relative to the planning. Cementless tibial component impaction resulted in the most deviation from plan, with a large proportion of cases resulting in being more prominent than planned.

Level Of Evidence: Level III.

Download full-text PDF

Source
http://dx.doi.org/10.1002/ksa.12171DOI Listing

Publication Analysis

Top Keywords

total knee
8
knee arthroplasty
8
robotic system
8
final components
8
small deviations
4
deviations planned
4
planned performed
4
performed bone
4
bone cuts
4
cuts ct-based
4

Similar Publications

Tissue Engineering Construct for Articular Cartilage Restoration with Stromal Cells from Synovium vs. Dental Pulp-A Pre-Clinical Study.

Pharmaceutics

December 2024

Sports Medicine Division, Institute of Orthopedics and Traumatology, Hospital das Clínicas HCFMUSP, Faculdade de Medicina, Universidade de São Paulo, São Paulo 05403-010, SP, Brazil.

Background/objectives: Cartilage injuries and osteoarthritis are prevalent public health problems, due to their disabling nature and economic impact. Mesenchymal stromal cells (MSCs) isolated from different tissues have the immunomodulatory capacity to regulate local joint environment. This translational study aims to compare cartilage restoration from MSCs from the synovial membrane (SM) and dental pulp (DP) by a tissue-engineered construct with Good Manufacturing Practices.

View Article and Find Full Text PDF

Assessment of Quadriceps Muscle in Advanced Knee Osteoarthritis and Correlation with Lower Limb Alignment.

Medicina (Kaunas)

December 2024

Department of Orthopedic Surgery, Keimyung University Dongsan Hospital, Keimyung University School of Medicine, Daegu 42601, Republic of Korea.

: Despite extensive studies of the role of quadriceps and quadriceps/hamstring balance in knee osteoarthritis (OA), the roles of the vastus intermedius, medialis, and lateralis in OA remain unclear. The purpose of this study was to investigate the relationship of lower limb alignment and the ratio of the quadriceps femoris muscle to the knee extensor muscle. : This study included 50 patients with advanced knee OA (Kellgren/Lawrence grade of 3 or 4) and 25 healthy control persons between June 2021 and May 2022.

View Article and Find Full Text PDF

The aim of this study was to test whether there is a necessity for routine postoperative laboratory testing in patients undergoing primary elective unilateral total knee arthroplasty (TKA), with the administration of Tranexamic Acid (TXA) and without the use of tourniquet. : This observational, retrospective cohort study was conducted at Meir Medical Center. The data were collected in 2018-2022.

View Article and Find Full Text PDF

: Abnormalities in patellar tracking, often overlooked in surgical planning, have been identified as a contributing factor to total knee arthroplasty (TKA) complications, including anterior knee pain, patellar subluxation, and dislocation. This study aims to evaluate the repeatability of a novel intraoperative setup for assessing patellofemoral kinematics and its interaction with prosthesis design and positioning during surgery. This setup may support personalized alignment techniques in TKA, potentially improving surgical outcomes.

View Article and Find Full Text PDF

The Etiology of Total Knee Arthroplasty Failure Influences on Improvement in Knee Function: A Follow-Up Study.

J Clin Med

December 2024

Department of Orthopedics, Klinik Penzing, Wiener Gesundheitsverbund, 1140 Vienna, Austria.

Outcomes following total knee arthroplasty (TKA) revisions are variable, and it is hypothesized that the underlying cause of primary TKA failure impacts postoperative outcomes. This study analyzes the results of TKA revisions seven years after surgery, in relation to the etiology of primary failure and other influencing factors. A previous study conducted in 2013 examined the same cohort of patients three months after revision surgery.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!